Maintenance - Optex SC-T4 Instruction Manual

Hide thumbs Also See for SC-T4:
Table of Contents

Advertisement

10

Maintenance

Maintenance
When the Cover or the Display is soiled, first wipe away lightly with a clean soft
cloth damped by diluted mild detergent solution and then wipe off moisture with a
dry clean soft cloth.
CAUTION:
Do not wipe the transmitter with solvent such as Thinner.
Periodic inspection
Inspect the following items every 3 months:
Make sure that the Transmitter is fixed securely.
Make sure that the Transmitter is not damaged.
Make sure that the screws of the Terminal Block are not rusted.
Calibration
The SC-T4 is designed so as to make measurements stably for a long time. In
order to maintain the reliability of measurement, however, carry out calibration at
least once a year (Refer to page 10 ([5] Operation (1) Calibration)).
Long-term storage
When the SC-T4 is not used for a prolonged period, turn off the Power Switch
and disconnect the Power Cable from the power supply prior to safekeeping.
11
Specifications
Model No.
Power supply
Power
Normal: 8VA or less, during washing: 16VA or less
consumption
(including detector, at the time of analog signal input/output of 20mA)
Display resolution
0.1mg/l (0-49.9mg/l), 1mg/l (50-1000mg/l)
Output
Signal Output (analog 4-20mA, resistance load of 300 or less)
Self-checking Relay Output (non-voltage C-contact capacity 240 VAC, 1A resistance load)
Alarm Relay Output (non-voltage C-contact capacity 240 VAC, 1A resistance load)
Detector signal input (analog 4-20mA, input resistance approx.100 )
Input (detector)
Operating temperature
-20 to +50
Main material
Dimensions
Weight
Approx. 1.6kg (including the Mounting Bracket)
Degree of protection
The specifications herein are subject to change without prior notice due to improvements.
12
Dimensions
Installation hole
16
Applicable cable
diameter
38
6~8
75
SC-T4
AC100
240V 10%
50/60Hz
Calibration Output
Power Supply for Sensor (12 VDC)
Self-checking input
, humidity 95%Rh or less (Avoid direct sunlight)
Polycarbonate
HxWxD = approx.162x240x75 mm
IEC 60529 / IP65 (jetproof type)
9
34
34
34
Hole for
Installation pitch 186
padlock
220
7
240
34
[mm]

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents