Signal Output Response Time Setting

Signal Output Response Time can be set for 1 to 120 seconds or OFF.
Setting unit is 1 second
The Signal Output Response Time is set for OFF as a factory setting.
Ex: To set the Signal Output Response Time for 10 seconds ;
1
Press
and select the
Response
Time.
2
Press
for 2 seconds or more, the
Signal Output Response Time
blinks.
3
Enter "10" on the Display.
Reference
Move the digit to be entered using the right and left arrow buttons
( 1 digit
10 digit
At this time, the selected digit blinks.
Enter a numerical value ( 0, 1, 2, ..., 9 ) into the selected digit using the
upper and lower arrow buttons
value and
button decreases the value.
While pressing
changed.
To cancel the Signal Output Response Time Setting, press
To change the Display for the Measurements Indication mode, press
again.
Reference:
To cut off the Signal Output Response Time, enter "0". Then "OFF"appears on the
Display. Next, press
.
4
Press
to terminate the Signal Output Response
Time Setting.
CAUTION:
When Err6 is displayed, the value is out of the setting range.
Signal Output
Indicator
100 digit ) .
.
button increases a numeric
buttons, the values displayed are continuously
3
Signal Output Range Setting
Signal Output Range can be set as follows.
"High Limit" "Low Limit"
Ex) when the setting value of Span-Adjustment is changed to
2000mg/l, the Signal Output Setting Range is as follows:
"High Limit" "Low Limit"
The Signal Output Range is default at 4mA for the 0-Adjustment
value and 20mA for the Span-Adjustment value.
Pip
CAUTION:
The resolution of the Analog Signal Output change depending on the setting value of
the Span-Adjustment.
Pip Pip
The resolution of Analog Signal Output is 0.5 mg/l as a factory setting. When it is
necessary to get 0.1mg/L as the resolution of Analog Signal Output, set the Span-
Adjustment to "1000mg/l" and the Signal Output Range to "200mg/l".
Ex: To set the The Signal Output Range for 100
( 1 ) Signal Output Low Limit Setting
1
Press
Output Low
"-Lo-" and "the setting value" comes on
alternatively.
2
Press
The Signal Output Low Limit value is
.
displayed.
3
Enter "100" on the Display.
Reference
Span-Adjustment value
400 (
and select the
The Signal
Limit.
for 2 seconds or more, the
Move the digit to be entered using the right and left arrow buttons
( 1 digit
100 digit ) .
10 digit
100 digit
At this time, the selected digit blinks.
Enter a numerical value ( 0, 1, 2, ..., 9 ) into the selected digit using the
upper and lower arrow buttons
value and
button decreases the value.
While pressing
buttons, the values displayed are continuously
changed.
To cancel the Signal Output Low Limit Setting, press
To change the Display for the Measurements Indication mode, press
again.
5
2000
5)
300 mg/l ;
Pip
Pip Pip
.
button increases a numeric
.