Analog programming and monitor

There are two MONITOR output BNC connections on the rear panel of the Model 2290-10. Both are 0
to +10 V outputs. There is also a SET BNC connection that is used to receive analog programming
voltage.
To select the rear SET analog high voltage control, press SELECT on the front panel until the REAR
LED is on, then press ENTER.
The next two figures show examples of analog input and output connections.
The BNC jack outer shells are connected to chassis ground and cannot be floated.
Figure 6: Model 2290-10 analog input control and voltage or current monitor connections
2290-10-900-01 Rev. A/December 2013
Section 3: Basic operations
3-5