Device-Specific
Commands
header could not be used because its value is outside the valid range
defined by the generator.
-224,"Illegal parameter value". A discrete parameter was received
which was not a valid choice for the command. An invalid parameter
choice may have been used.
-300,"Device-specific-error". This is the generic device-dependent
error for the instrument when it cannot detect more specific errors. A
device- specific error as defined in IEEE-488.2 has occurred.
-311,"Memory error". Indicates that an error was detected in the
instrument s memory.
-350,"Too many errors". The error queue is full because more than
30 errors have occurred. No additional errors are stored until the
errors from the queue are removed. The error queue is cleared
when power has been shut off, or after a *CLS command has been
executed.
-410,"Query INTERRUPTED". A command was received which
sends data to the output buffer, but the output buffer contained data
from a previous command (the previous data is not overwritten). The
output buffer is cleared when power is shut off or after a device clear
has been executed.
The Model 3151/3151A conforms to the 1993.0 version of the SCPI
standard. Some of the commands used are not included in the
1993.0 version. However, these commands are designed with the
SCPI standard in mind and they follow all of the command syntax
rules defined by the standard. Table 4-1 lists all device-specific
commands that were designed specifically for the Model 3151/3151A
as not confirmed SCPI 1993.0 commands.
SCPI Command Reference 4-23