Recommended Test Equipment; Performance Check Procedures - Racal Instruments 3151 User Manual

Waveform generator
Table of Contents

Advertisement

3151 And 3151A User Manual
Recommended Test
Equipment
Test Instrument & Accessories
Universal Counter
DMM
Distortion Analyser
Spectrum Analyser
Pulse/Function Generator
Oscilloscope
Feedthrough termination
BNC Cables
20dB feedthrough attenuator
Performance Check
Procedures
Recommended test equipment for troubleshooting, calibration and
performance checking is listed below. Test instruments other than
those listed may be used only if their specifications equal or exceed
the required characteristics.
Minimum Specification
100 MHz, 10 ppm accuracy
10 Hz to 100 kHz
1 GHz bandwidth, 1 kHz resolution
20 MHz, 10 ns pulsewidth
50Ω, 1%
Use the following procedures to check the Model 3151/3151A
against the specifications. A complete set of specifications is listed
in Appendix A. The following paragraphs show how to set up the
instrument for the test, what the specifications for the tested function
are, and what acceptable limits for the test are. If the instrument
fails to perform within the specified limits, the instrument has to be
calibrated or tested to find the source of the problem.
ACV, 0.2%; DCV, 0.1% accuracy
2 channels, 400 MHz analog bandwidth
1 meter long, 50Ω impedence
50Ω, 1%
Maintenance and Performance Checks 5-5