Amplitude Accuracy - Racal Instruments 3151 User Manual

Waveform generator
Table of Contents

Advertisement

3151 And 3151A User Manual

Amplitude Accuracy

Accuracy Specifications (1 kHz):
"(1% of reading +200 µV), to 160 mV
"(1% of reading +2 mV), to 1.6 V
"(1% of reading +20 mV), to 16 V
Equipment: DMM, 50Ω feedthrough termination
1. Configure the Model 3151/3151A as follows:
Function:
Frequency:
Amplitude:
Remote Commands:
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15
2. Set the DMM to ACV measurements (RMS).
3. Connect the Model 3151/3151A output to DMM input. Terminate
the output with a 50Ω feedthrough termination. Set the
amplitude and verify DMM reading as follows:
3151 SETTING
15.0 V
1.50 V
150 mV
Sine
1.000 kHz
15 V
:RES
DMM READING
5.373 V - 5.230 V
537.3 mV - 523.0 mV
53.73 mV - 52.30 mV
Maintenance and Performance Checks 5-7