Racal Instruments 3151 User Manual page 111

Waveform generator
Table of Contents

Advertisement

3151 And 3151A User Manual
Burst
1. Configure the Model 3151/3151A as follows:
Function:
Frequency:
Operating Mode:
Remote Commands:
2. Set the oscilloscope and verify that the Model 3151/3151A
outputs a burst of 10 complete output waveforms.
Remove the external pulse generator connection from
the Model 3151/3151A for the next test.
Sine wave
1.000 kHz
Burst
:RES
:OUTP ON
:FREQ 1e3
:INIT:CONT OFF
:TRIG:BURST:STAT ON
:TRIG:COUN 10
:TRIG:SOUR:ADV:EXT
Maintenance and Performance Checks 5-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

3151a

Table of Contents