Re-Calibration Mode

Step-by-step Instructions
for RE-CAL Mode
16
The re-calibration (RE-CAL) mode menu lets you
change the scale resolution or rounding method
without using any calibration weights. If you want
to change the unit of measure operation, you
must perform a full calibration using test weights.
For a scale originally calibrated in the lb. mode,
you may also change rounding methods (i.e.,
scale or classifier).
Follow these steps to re-configure your scale
(without weights). Refer to Figure 2.
1. From the normal weighing mode, press SW-
1...
2. Press the ZERO key until...
3. Press the TEST key...
4. Press the ZERO key until desired capacity/
resolution is displayed.
5. Press the TEST key to select a new capacity/
resolution. If the scale is operating in LB
mode...
6. Press the ZERO key to toggle between
SCALE and CLASS.
7. When the choice you want is displayed,
press the TEST key...
8. Press the TEST key or SW-1 to return to
normal weighing mode.
Model 7815 Parcel Bench Scale User's Manual
DIAG is displayed.
RE-CAL is displayed.
150.1 is displayed. (Displays the current
capacity/resolution setting)
The unit will display CLASS.
DONE is displayed.