Transceiver And Retro Reflector Assembly - EMS 1304 Installation And Operation Manual

Opacity monitor
Table of Contents

Advertisement

Transceiver and retro reflector assembly

Attach the Transceiver and Retro reflector to
1.
the air plenum assembly by placing them on
the hinge pins.
Close transceiver & retro and secure in place
2.
with the two hold down latches.
The air-purge blowers should be powered up
3.
at this time to prevent stack particulate from
accumulating in the nipple and air-purge
housing.
Caution: If installed location has a positive pressure
the air-purge system must be used continuously during installation to prevent process gases from contaminating
optical surfaces or over heating instrument electronics. If the system is shut off for more than momentary
interruptions, the instrument may be damaged. Failure to provide continuous air-purge may void the warranty.
All wiring from the control unit to the transceiver should be completed at this time.
NOTE: THE AIR PLENUM ASSEMBLY FOR BOTH TRANSCEIVER AND RETRO MUST BE
INSTALLED AS BELOW, I.E. THE PINS ON THE LEFT SIDE POINTING UP!
Lift side Upper
Hinge Pin
Left side Lower
Hinge Pin
19