1304 Opacity Specifications - EMS 1304 Installation And Operation Manual

Opacity monitor
Table of Contents

Advertisement

1304 OPACITY SPECIFICATIONS

Design and performance:
Peak and Mean Spectral Response
Relative spectral response
Angle of View
Angle of Projection
Calibration Error/accuracy
Response time
Zero Drift
Calibration Drift
Operational Period
Zero/Span Calibration
Remote control unit:
Numeric Digital Display 5/8 inch (15.9 mm), 4-
digit LED
Display resolution
Alphanumeric 8 character LED (0.2") Fault Display Air purge low, No Stack Power, T2 4-20 Lost/low, Maintenance Mode, Window Dust, Zero Cal
Bargraph
Operator Indicating LED's
High Opacity Set Point
High Opacity Delay
Hi Opacity Alarm Reset
Open Transistor Collector Outputs
Dry Contacts
Instantaneous Opacity and Average Outputs
OPLR (Exit Correlation Lx / 2*Lt)
Cal cycle initiate
Enclosure
Ambient Temperature Range
Power Requirements
Network
Configuration
Transceiver Service module:
Display
Local zero/span
Test jacks
Diagnostics
Transceiver/ Reflector:
Meets or Exceeds PS-1 Revised and ASTM D6216-03
Photopic; 515 to 585 nm, less than 10% of peak response outside 400 to 700 nm
<10%
< 4.0
from optical axis
o
< 4.0
from optical axis
o
+/- 1% of full scale
< 10 second
< 0.5%
< 0.5%
In excess of PS-1 required 336 Hrs.
Manual or automatic with zero mirror and neutral density filter
120Vac +/- 10% 50/60hz 22va
Selectable for Instantaneous opacity, Average Opacity. Optional mg/m
0.1% Opacity
Fail, Span Cal Fail, Manual Cal.
Left/Right: 51 segments LED (5.1") for instant/average opacity. 20 segment LED (2.1") for
window dust and drift in %opacity
Run, Alarm, Operate, Cal in progress
0-100% of full scale
Adjustable 0-300 Seconds
Automatic
Four selectable
Two selectable (option: two additional)
4-20 mA grounded reference, 800 ohms max. (two additional optional)
0.2 to 3.0
Local, remote and timed Adjustable 1 to 24 hours
NEMA 13, IP65, Height: 7.38" x Wide: 3.00: x Depth: 7.75", Weight: 6 lbs.
+0
to +50
C ( +32
to + 122
o
o
o
120 VAC (+/- 15%), 50/60 Hz, 22VA
Protocol: Modbus (ASCII or RTU mode), type RS-485, optically isolated.
Speed: 300-38,400 baud, type RS-232
120/220Vac +/- 10% 50/60hz, Transceiver 65va, Service Module 60va
5/8 inch (15.9 mm), 4-digit LED, selectable for % Opacity and % Transmittance
Manual on demand
Transceiver to remote control current loop
Loss of power, Current loop open, maintenance mode.
12
F)
o
3