Chapter 8: Troubleshooting
PROBLEM
Replace filters
intermittent alarm
occurs
Contaminants outside
of enclosure
Blower will not
operate
Low face velocity
Blower and lights will
not operate
60
CAUSE
Organic sensor alarms
intermittently and display
reads "Replace Filters-s",
s = organic sensor
Carbon filter loaded with
chemicals.
Improper user techniques
for the enclosure.
Restriction of the baffle
air slots or blockage of
the exhaust outlet.
External factors are
disrupting the enclosure
airflow patterns or acting
as a source of
contamination.
Enclosure has improper
face velocity.
Unit not plugged into
outlet.
Circuit breaker(s) or
Ground Fault Interrupter.
Blower wiring is
disconnected.
Blower switch is
defective.
Motorized impeller is
defective.
Enclosure sash not closed. Close sash to the lowest position.
HEPA filter clogged.
Unit not plugged into
outlet.
Circuit breaker(s) tripped.
CORRECTIVE ACTION
Organic sensor has detected filter
breakthrough and only 5-10% of filter life
remains. Replace filters with a spare set of
filters stored. Consult Chapters 3, 4, and 5,
Filter Specialist, and Chemical Guide for
filter life based on specific chemical use. All
chemicals have different life. Always limit
chemical use, cover up chemicals and turn
off blower when not in used to preserve life.
Replace carbon filter.
See "Validate the Enclosure" Chapter 3 and
"Safety Precautions" Chapter 4 sections in
the manual.
Remove obstruction to ensure that all air
slots and the exhaust outlet are unobstructed.
See "Location Requirements" Chapter 2,
"Validate the Enclosure" Chapter 3, and
"Safety Precautions" Chapter 4 sections of
this manual.
Have enclosure certified per Chapter 6 and
check exhaust system. Check carbon filters
and replace if necessary. Enclosure should
have an average face velocity of 60-100 fpm.
Plug the enclosure into appropriate electrical
service.
Reset circuit breaker.
Inspect blower wiring.
Replace membrane switch on control panel.
Replace motorized impeller. See Chapter 6.
Replace HEPA filter.
Plug enclosure into appropriate electrical
service.
Reset or replace circuit breaker.
Product Service 1-800-522-7658