Calibration Check; Calibration Adjustment - GE K0341 User Manual

Intrinsically safe loop calibrator
Table of Contents

Advertisement

Applied V
0
20
40
50
UPS-III-IS source mode
Applied
mA
0
4
12
20

Calibration Check

1.Connect the UPS-III-IS to the electrical calibrator. Switch on
the electrical calibrator and allow it to thermally stabilise.
2.Switch on the UPS-III-IS and allow the instrument to thermally
stabilise.
3.Set the UPS-III-IS to mA measure, adjust the electrical
calibrator to apply the first value in the table 3. Record the
reading of the UPS-III-IS.
4.Repeat step 3 for all the values in the table 3.
5.Compare the recorded values and the applied values if the
difference is greater than the permitted error, the instrument
requires a calibration adjustment.
6.Repeat this procedure for V measure (table 4) and mA source
(table 5).

Calibration Adjustment

1.Connect the UPS-III-IS to the electrical calibrator. Switch on
the electrical calibrator and allow it to thermally stabilise.
K0341 Issue No. 2
Table 4
V measure
Permitted UPS-III-IS
error (mV)
0.004
0.004
0.005
0.005
Table 5
mA source
Permitted UPS-III-IS
error (mA)
0.002
0.002
0.002
0.002
Calibrator error
Calibrator error
14
(mV)
0.00040
0.00014
0.00064
0.00070
(mA)
0
0.00012
0.00011
0.00015