GE K0341 User Manual page 9

Intrinsically safe loop calibrator
Table of Contents

Advertisement

Guide to supply and input/output
No.
Current measurement between mA and COM with
1
external 24V. This mode of operation inserts the apparatus
&
in the current loop by breaking into the circuit and
2
connecting mA (positive) and COM (negative) into the
circuit.
Current measurement between mA (24V) and mA with
3
internal 24V. Terminal mA(24V) provides the source of
power to supply remote sensor.
Special condition for safe use - remote sensor MUST be
isolated from all other sources of power.
Voltage measurement between V and COM
4 4
Continuity measurement between V and COM
5
Special condition for safe use - remote sensor MUST be
isolated from all other sources of power.
Current generation between mA and COM with external
6
24V. This mode of operation inserts the apparatus in the
current loop by breaking into the circuit and connecting
mA (positive) and COM (negative) into the circuit.
Current generation between mA (24V) and mA with
7
internal 24V. Terminal mA (24V) provides the source of
power to supply the remote sensor.
Special condition for safe use - remote sensor MUST be
isolated from all other sources of power.
Table 1
Parameter and conditions
3
K0341 Issue No. 2