YSI 6 series User Manual page 36

6-series multiparameter water quality sondes
Table of Contents

Advertisement

Sondes
ENABLING SENSORS
To activate the sensors that are in your sonde, select Sensor from the Sonde Main menu.
------------Sensors enabled------------
1-(*)Time
2-(*)Temperature
3-(*)Conductivity
4-(*)Dissolved Oxy
5-(*)ISE1 pH
6-(*)ISE2 Orp
7-(*)ISE3 NH4+
8-(*)ISE4 NO3-
9-( )ISE5 NONE
A-(*)Optic T Turbidity - 6136
B-(*)Optic C Chlorophyll
Select option (0 for previous menu):
Note that the exact appearance of this menu will vary depending upon the sensors that are available on your
sonde. Enter the corresponding number to enable the sensors that are installed on your sonde. An asterisk
indicates that the sensor is enabled.
When selecting any of the ISE or Optical ports, a submenu will appear. When this occurs, make a selection
so that the sensor corresponds to the port in which the sensor is physically installed. Only ORP can be
enabled as ISE2. Optic T, Optic C, Optic B, and Optic O generate a submenu on selection. Each optical
port can have one of six probes (6136 Turbidity, 6025 Chlorophyll, 6130 Rhodamine WT, 6131 BGA-PC,
6132 BGA-PE, or 6150 ROX Optical DO) installed as indicated by the submenus.
NOTE CAREFULLY: It is NOT possible to simultaneously activate BOTH the 6562 Rapid Pulse
polarographic dissolved oxygen sensor and the 6150 ROX Optical dissolved oxygen sensor. Activation of
either sensor will automatically deactivate the other selection. Thus, users of 6600V2-2, 6600EDS V2-2,
6820V2-1, and 6920V2-1 sondes CANNOT measure oxygen with both types of sensors.
After all installed sensors have been enabled, press Esc or 0 to return to the Main Menu.
ENABLING PARAMETERS
In order for a specific parameter to be displayed:
1.
The sensor must first be enabled as described above.
2.
That parameter must be activated in the Report Setup menu described below.
Select Report from the Main menu. A Report Setup menu similar to the one shown below will be
displayed.
YSI Incorporated
Environmental Monitoring Systems Operations Manual
Section 2
2-30

Hide quick links:

Advertisement

Table of Contents
loading