YSI 6 series User Manual page 312

6-series multiparameter water quality sondes
Table of Contents

Advertisement

Solubility and Pressure/Altitude Tables
Table 2: Calibration Values for Various Atmospheric Pressures and Altitudes
PRESSURE
Inches Hg
mm Hg
30.23
768
29.92
760
29.61
752
29.33
745
29.02
737
28.74
730
28.43
722
28.11
714
27.83
707
27.52
699
27.24
692
26.93
684
26.61
676
26.34
669
26.02
661
25.75
654
25.43
646
25.12
638
24.84
631
24.53
623
24.25
616
23.94
608
23.62
600
23.35
593
23.03
585
22.76
578
22.44
570
22.13
562
21.85
555
21.54
547
21.26
540
20.94
532
20.63
524
20.35
517
20.04
509
19.76
502
YSI Incorporated
Environmental Monitoring Systems Operations Manual
ALTITUDE
Millibars
Feet
1023
-276
1013
0
1003
278
993
558
983
841
973
1126
963
1413
952
1703
942
1995
932
2290
922
2587
912
2887
902
3190
892
3496
882
3804
871
4115
861
4430
851
4747
841
5067
831
5391
821
5717
811
6047
800
6381
790
6717
780
7058
770
7401
760
7749
750
8100
740
8455
730
8815
719
9178
709
9545
699
9917
689
10293
679
10673
669
11058
CALIBRATION
VALUE
Meters
Percent Saturation
-84
101
0
100
85
99
170
98
256
97
343
96
431
95
519
94
608
93
698
92
789
91
880
90
972
89
1066
88
1160
87
1254
86
1350
85
1447
84
1544
83
1643
82
1743
81
1843
80
1945
79
2047
78
2151
77
2256
76
2362
75
2469
74
2577
73
2687
72
2797
71
2909
70
3023
69
3137
68
3253
67
3371
66
Appendix D
D-
3

Hide quick links:

Advertisement

Table of Contents
loading