Product Test; How To Purge The Product Sensor; How To Take A Sample - Fluke 985 Getting Started Manual

Airborne particle counter
Hide thumbs Also See for 985:
Table of Contents

Advertisement

985
Getting Started

Product Test

Electrical noise, sensor leakage, or other interference can
cause the Product to give incorrect data.
To check operation:
1.
Attach the Zero-Count Filter.
2.
In the Data Display Setup screen, enable the 0.3 μm
channel and set Concentration mode to COUNTS.
3.
In the Sample Setup screen:
1.
Set the sample time to 5 minutes
2.
Set the hold time to 00:00:00
3.
Set the delay time to 00:00:03
4.
Set the MODE to Automatic
5.
Set the number of CYCLES = 2.
4.
Start the Product sampling and let it complete the 2 x
5 minute samples.
5.
Examine the particle counts from the sample. Counts
must align with these specifications for instrument
operation to be verified:
No more than 1 particle >0.3 μm in 5 minutes
8

How to Purge the Product Sensor

Particles must be purged (removed) from the Product
sensor before use in a clean room or clean environment,
and after each high-sampling count. To purge the sensor:
1.
Attach the adapter and the Zero-Count Filter.
2.
Set the Product to sample continuously (set the
Count Mode to Cumulative.)
3.
Start the count procedure. Continue the count
procedure until there are no new counts.
4.
Remove the Zero-Count Filter for normal operation.

How to Take a Sample

Before you take a sample, make sure that you remove
the protective cap. To take a sample, select the Sample
screen and push . The Product samples with values in
the Sample Setup menu. To stop a sample, let the
sample procedure complete or push  again. See the
"Sample Setup Menu" section to set the sample values.