Input Impedance
Function
Volts, mV
Common Mode Rejection Ratio
DC Volts, mV
>100 dB at dc, 50 Hz or 60 Hz 0.1% >90 dB at 50 Hz or 60 Hz 0.1%
AC Volts, mV
> 90 dB dc to 60 Hz
Open Circuit Test Voltage
Ohms
< 5V
Diode Test
< 5V
500
Ohms
1 A
Diode Test

Safety Information

Function
Description
Safety
1000 V, AC/DC, maximum voltage between any
terminal and earth ground.
Complies with ANSI/ISA-S82.01-94, CSA C22.2
No 1010.1-92 to 1000 V Overvoltage Category III
and to 600V Overvoltage Category IV.
Certification agencies (approvals/listings pending):
UL
per standard UL 3111 (pending)
CSA per standard CSA/CAN C22.2 No. 1010.1-92
TÜV per standard EN 61010 Part 1-1993 (pending)
Surge Protection
8 kV peak per IEC 1010.1-92
Fuse Protection
440 mA, 1000 V FAST Fuse, 11A, 1000V FAST Fuse
Markings
UL, CSA, TUV, CE (VDE pending)
Input Impedance (Nominal)
10 M , < 100 pF
Normal Mode Rejection
Full-Scale Voltage
To 5 M
500 mV
3.1000V
Typical Short-Circuit Current
5 k
50 k
500 k
100 A
10 A
1 A
1.0 mA typical
30 to 500 M + nS
3.1V
5 M
30 M
0.1 A
0.1 A

General Specifications

Function
Description
Display
Digital: 50000/5000 counts primary, 5000 counts secondary,
updates 4/second.
Analog: 51 segments, updates 40/second.
Operating Temperature
-20 C to +55 C
Storage Temperature
-40 C to +60 C
Temperature Coefficient 0.05 x (specified accuracy)/ C (<18 C or >28 C)
Relative Humidity
0% to 90% (0 C to 55 C)
0% to 70% (35 C to 55 C)
Altitude
Operating: 2000 meters
Storage: 10000 meters
Battery Type
4 AA Alkaline, NEDA 15A or LR6
Battery Life
72 Hours typical (with backlight off)
Shock Vibration
Per MIL-T-PRF 28800 for Class II instruments
EMC
Susceptibility and Emissions: Commercial Limits per EN61326-1
Size
10.0 cm X 20.3 cm X 5.0 cm (3.94" X 8.00" X 1.97")
(Not Including Accessory Mount)
Weight
545 grams (1.2 lbs.)
Case Sealing
IP-42 per IEC 529, Section 3
Warranty
Lifetime
Calibration Interval
1 year