187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
Detailed Specifications
Accuracy is specified for a period of one year after calibration, at 18 C to
28 C (64 F to 82 F) with relative humidity to 90%.
Accuracy specifications are given as ([% of reading] + [number of least
significant digits])
Function
Range
Resolution
1
Resistance
500.00
0.01
5.0000 k
0.0001 k
50.000 k
0.001 k
500.00 k
0.01 k
5.0000 M
0.0001 M
30.000 M
0.001 M
100.0 M
0.1 M
500.0 M
0.1 M
Conductance
50.00 nS
0.01 nS
3
Capacitance
1.000 nF
0.001 nF
10.00 nF
0.01 nF
100.0 nF
0.1 nF
1.000 F
0.001 F
10.00 F
0.01 F
100.0 F
0.1 F
1.000 F
1 F
10.0 mF
0.01 mF
50.00 mF
0.01 mF
Diode Test
1
3.1000V
0.0001V
Frequency
500.00 Hz
0.01 Hz
5.0000 kHz
0.0001 kHz
50.000 kHz
0.001 kHz
999.99 kHz
0.01 kHz
Duty Cycle
10.00 to 90.00% 0.1%
5
Pulse Width
499.99 ms
0.01 ms
999.9 ms
0.1 ms
Temperature
-200 to
0.1 C
+1350 C
-328 to
0.1 F
+2462 F
Min-Max-Avg Response: 100 ms to 80%
8
Fast Min-Max 250 s
1
For the 5,000 count mode, divide the number of least significant digits
(counts) by 10.
2
For relative humidity greater than 70%, resistance accuracy is 0.5%
over 1 M and 2.5% over 10 M .
3
For film capacitor or better, using Relative mode (REL ) to zero
residual on 1.1 nF range.
4
Reading will be 0.00 for signals below 0.5 Hz.
5
Least significant digit not active above 10 mF.
6
Using relative mode (REL ) to zero residual reading.
7
For ambient temperature changes of
1 hour.
8
For repetitive peaks, 2.5 ms for single events.
9
Frequency greater than 5 Hz, except for VDC, 500 mV dc and
3000 mV dc functions; 0.5 Hz to 1.0 + 0.
10
Range/input ratios also apply to current functions.
11
To ensure stated accuracy, switch to conductance mode and verify that
the open circuit reading is less than 0.10 nS.
12
Accuracy specification is relative to the user - adjustable temperature
offset, and assumes ambient temperature stable to
Above specifications are subject to change without notice.
(continued)
Accuracy
6
0.05% + 10
0.05% + 2
0.05% + 2
0.05% + 2
0.15% + 4
2
1% + 4
2
3% + 2
11
10% + 2
11
1% + 10
2% + 5
1% + 5
5
3% + 10
2% + 20
4
(0.005% + 1)
(Voltage Range/
Input Voltage) x 300
9,10
counts)
(3% x(Voltage range/
input voltage)+ 1 count)
9,10
7,12
(1% of reading + 1 C)
(1% of reading +
1.8 F)
7,12
Specifed accuracy 12
counts for changes
>200 ms in duration.
( 40 digits in AC for
changes >350 ms and
inputs >25% of range)
Specified accuracy 100
counts for changes
8
>250 s in duration
5 C, rated accuracy applies after
1 C
Memory and PC Communication
Functions
(Fluke 189 Only)
The Fluke 189 adds the following capabilities:
Interval LOGGING
At least 288 intervals (specified by user in Setup)
may be recorded to internal memory. These
values may be viewed using the VIEW MEM
function on the meter. Up to 700 unstable event
values (similar to AutoHold) are automatically
added to LOGGING memory for viewing only
through the optional FlukeView
software. Additional intervals will be logged if
the signal is stable.
Reading SAVE
Up to 100 readings may be saved by the user in a
memory separate from LOGGING memory. These
readings may be viewed using VIEW MEM.
Frequency Counter Sensitivity
Approximate VAC Sensitivity
(RMS Sine Wave)
1
15 Hz to
500 Hz
Input Range
100 kHz
2
50 mV
5 mV
10 mV
500 mV
20 mV
20 mV
3000 mV
500 mV
2000 mV
5V
0.5V
2.0V
50V
5V
5.0V
500V
20V
20V
1000V
100V
100V
1
Maximum imput = 10 x Range (1000 V max). Noise at low frequencies and amplitudes may
affect accuracy.
2
Usable at reduced sensitivity to 0.5 Hz and 1000 kHz.
3
Typical frequency bandwidth with full scale (or maximum 2 x 10
RMS sine wave.
Burden Voltage (A, mA, A)
Function
Range
mA- A
500.00 A
5,000 A
50.000 mA
400.00 mA
A
5.0000 A
10.000 A
®
Forms PC
2
VAC
Approximate VCD
Bandwidth
3
Trigger Levels
1
1 MHz
-5 mV & 5 mV
1 MHz
5 mV & 65 mV
800 kHz
140 mV & 200 mV
950 kHz
1.4 V & 2.0 V
1 MHz
0.5 V & 6.5 V
1 MHz
5 V & 65 V
> 400 kHz
5 V & 65 V
7
V-Hz product)
Burden Voltage (typical)
102 V / A
102 V / A
1.8 mV / mA
1.8 mV / mA
0.04 V / A
0.04 V / A
VDC
Bandwidth
3
1 MHz
1 MHz
90 kHz
14 kHz
> 400 kHz
> 400 kHz
> 400 kHz