Fluke 113 digital multimeter datasheet

Advertisement

Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C,
with relative humidity at 0 % to 95 %. Accuracy specifications are given as follows:
Note:
Specifications are subject to change without notice.
± ( [ % of reading ] + [ number of least significant digits ] )
Function
Range
Resolution
6.000 V
0.001 V
1,2
Chek
60.00 V
0.01 V
600.0 V
0.1 V
Range
Resolution
Function
2.000 V
0.001 V
3
3
600.0 Ω
0.1 Ω
3
6.000 kΩ
0.001 kΩ
60.00 kΩ
0.01 kΩ
1000 nF
1 nF
3
10.00 µF
0.01 µF
100.0 µF
0.1 µF
9999 µF
1 µF
Notes:
1
All
Chek voltage ranges are specified from 60 counts to 100 % of range. Because inputs below 60 counts are not
specified, it is possible and normal for this and other true-rms meters to display non-zero readings when the test
leads are disconnected from a circuit or are shorted together.
2
Crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale.
3
After measuring voltage, a wait time of 1 minute is required to maintain accuracy of ohms, capacitance, diode test,
and continuity.
Maximum voltage between any
terminal and earth ground
Display
Operating temperature
Storage temperature
Temperature coefficient
Operating altitude
Storage altitude
Relative humidity
Battery type
Battery life
Shock
Vibration
Size (HxWxL)
Weight
Safety
EMI regulations
Function
Input Impedance (Nominal)
Chek
Common Mode Rejection Ratio (1 kΩ Unbalanced)
> 60 dB at dc, 50 or 60 Hz
Open Circuit Test Voltage
< 2.7 V dc
< 2.7 V dc
2 Fluke Corporation
113 Basic Meter
Accuracy
DC, 45 to 500 Hz
2.0 % + 3
Accuracy
2.0 % + 3
Beeper on < 20 Ω, off > 250 Ω; detects
opens or shorts of 500 µs or longer
0.9 % + 2
0.9 % + 1
0.9 % + 1
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF - 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
600 V
3 3/4-digits, 6000 counts, updates 4/sec
-10 °C to 50 °C (14 °F to 122 °F)
-40 °C to 60 °C (-22 °F to 140 °F)
0.1 x (specified accuracy)/°C (< 18 °C or > 28 °C)
2,000 m
10,000 m
95 % to 30 °C
75 % to 40 °C
45 % to 50 °C
9 Volt Alkaline, ANSI 1604A / IEC 6F22
Alkaline: 300 hours typical, without backlight
1 Meter drop per IEC 61010-1-2001
Per MIL-PRF-28800 for Class 2 instrument
167.1 mm X 85.1 mm X 46.0 mm (6.58 in X 3.35 in X 1.81 in)
13.0 oz (404 g)
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No
61010-1-04, UL 61010-1 (2004) and IEC/EN 61010-1 2nd Edition for
measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Complies with FCC Part 15, Class B
~3 kΩ < 300 pF
Full Scale Voltage
< 0.7 V dc
2.000 V dc
Short Circuit Current
< 350 µA
< 1.0 mA
500 Hz to 1 kHz
4.0 % + 3
Ordering
information
Fluke 113 Basic Meter
Included
Fluke 113 Basic Multimeter
TL75 Test Leads
One 9V Battery
Printed Users Manual
Fluke. Keeping your world
up and running.
®
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2008 Fluke Corporation. Specifications subject
to change without notice. Printed in U.S.A.
3/2008 3292015 D-EN-N Rev A