Glitch Trigger - LeCroy 6000 SERIES Operator's Manual

Waverunner
Table of Contents

Advertisement

Operator's Manual
field. Or touch the GreaterThan button and enter a pulse-width value in the Lower Limit
data entry field. Or touch the InRange button. Touch the Delta button
nominal range, plus or minus a delta value in seconds. Touch inside the Nominal Width
and Delta data entry fields and enter values using the pop-up numeric keypads.
Alternatively, touch the Limits button
Touch inside the Lower Limit and Upper Limit data entry fields and enter values using
the pop-up keypads. Or touch the OutOfRange button and perform the same range
setups as for InRange triggering.
G
T
LITCH
RIGGER
How Glitch Trigger Works
Glitch trigger can be used to catch glitches. You can specify a pulse width or a pulse width range.
Pulse smaller than selected pulse width: Set a maximum pulse width. This glitch trigger is
generated on the selected edge (positive or negative) when the pulse width is less than or equal
to the set width.
The timing for the width is initialized and restarted on the opposite slope to that selected. You can
set widths from 600 ps to 20 s.
NOTE: If the glitch's width is narrower than the signal's width, set the trigger to a narrower width
than that of the signal. The signal's width, as determined by the instrument trigger comparator,
depends on the DC trigger level. If that level were to be set at the middle of a sine wave, for
example, the width could then be considered as the half period. But if the level were higher, the
signal's width would be considered to be less than the half period.
90
to set up a precise pulse-width range.
ISSUED: October 2003
to set up a
WR6K-OM-E Rev A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Waverunner 6000 series

Table of Contents