Battery powered, hand held instrument with a digital
lcd display capable of measuring the velocity, volume flowrate and temperature of air by means of a thermal probe (4 pages)
Page 1
COMBINED SERVICE MANUAL for: DIOXOR ® MONOXOR ® MONOXOR II H (High Range) ® NONOXOR ® Rev. 7 – May 2010 Bacharach, Inc. 621 Hunt Valley Circle, New Kensington, PA 15068 Phone: 1-800-736-4666 • Fax: 724-334-5001 • Web: www.mybacharach.com Printed in the U,S.A.
Sensor Replacement................9 Printed Circuit Board Replacement.............10 Chamber Assembly Servicing..............11 SENSOR CALIBRATION................12 Monoxer II CO Calibration..............12 Dioxor II SO2, Monoxor II H CO, and Nonoxor II NOx Calibration...13 TROUBLESHOOTING CHART ..............14 REPLACEMENT PARTS................17 Probe Assembly .................17 Instrument Assembly ................18 Miscellaneous Parts and Accessories..........20...
Single Gas Instrument Service Manual SERVICE MANUAL CONTENTS This manual contains the following service information for the . . . DIOXOR II (Part No. 0019-7038 & 0019-7043) MONOXOR II (Part No. 0019-7034 & 0019-7039) MONOXOR II H (Part No. 0019-8015) NONOXOR II (Part No.
(Chamber Assembly) and the Printed Circuit Board (P.C.B.) from the front case by pulling them straight up from the front case (Fig 2 -2); then lay these items on your workbench. Figure 2-1. Opening and Closing the Case Page 5 Instruction 0019-9166 Bacharach, Inc.
5. Reinstall the batteries (observe proper battery polarity); replace the battery cover; and attach the elastic strap’s metal clip to the bottom of the instru- ment. Figure 2-2. P.C.B. and Chamber Assem- Removal and Installation Page 6 Bacharach, Inc. Instruction 0019-9166...
6. Clean the Chamber’s inlet and outlet ports using pipe cleaners dipped in warm, soapy water. After cleaning, dry them with clean dry pipe cleaners. 7. Reassemble the chamber assembly as follows: NOTE: DO NOT overtighten any screws; damage to plastic parts will result. Page 7 Instruction 0019-9166 Bacharach, Inc.
Page 8
Step 2 (Tighten to 4–8 oz. in.). 9. Close the case as per Section “2.2 Closing the Case.” Figure 2-3. Chamber As- sembly, Exploded View Figure 2-3A. Fan-blade assem- bly set screw Page 8 Bacharach, Inc. Instruction 0019-9166...
Nos.): • Phillips-head screwdriver, No. 1 • Replacement Sensor: • Nutdriver, 4mm (5/32") • Dioxor II - SO • Screwdriver, 1/8" fl at blade • Monoxor II - CO • O-Ring and Membrane, if required • Nonoxor II - NO •...
7. Close the case as per Section 2.2 Closing the Case.” 8. Calibrate the Sensor with the new P.C.B. as per Section “3.0 Sensor Calibra- tion.” 9. Discard the old P.C.B. Figure 2-4. Sensor & Motor Wiring Diagram Page 10 Bacharach, Inc. Instruction 0019-9166...
6. Reassemble the chamber assembly with any replacement parts using Section 2.3, Step 7. 7. Reinstall the Membrane, O-ring, and Sensor using Section 2.3, Step 8. 8. Close the case as per Section “2.2 Closing the Case.” Page 11 Instruction 0019-9166 Bacharach, Inc.
NOTE: An indication that the instrument was not zeroed in fresh air is when a large negative CO display appears when the instrument is moved into another area. b. Set front panel POWER switch to ON. Figure 3-1. Calibration Kit Setup Page 12 Bacharach, Inc. Instruction 0019-9166...
CO gas in larger amounts, can be very toxic. Calibration at a Bacharach Sales/Service Center is recommended (See Section “5.4 Bacharach Sales/ Service Centers” to fi nd the nearest center). If you have a facility that is properly equipped to handle toxic gas (i.e..
Board Replacement.” LCD segments missing LCD loose or defective Carefully re-seat LCD in its socket. If segments are still missing, replace LCD. P.C.B. defective Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.” Page 14 Bacharach, Inc. Instruction 0019-9166...
Page 15
Sensor connections loose or Check Sensor connections. broken See Fig 2-2. Sensor depleted or defective Replace Sensor. Refer to Section “2.4 Sensor Replace- ment.” P.C.B. defective Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.” Page 15 Instruction 0019-9166 Bacharach, Inc.
Page 16
Check operation of Motor. If Motor defective motor is receiving power but isn’t turning, replace Motor Assembly. Refer to Section “2.6 Chamber Assembly Servicing.” Replace P.C.B. Refer to Sec- P.C.B. defective tion “2.5 Printed Circuit Board Replacement.” Page 16 Bacharach, Inc. Instruction 0019-9166...
Adapter, Motor Shaft 0024-0412 Battery Clip, push on 0004-3113 Battery Cover with Foam Pad 0019-3029 Chamber 0019-3090 Front Cover w/Display Window (Dioxor II) 0019-3079 Front Cover w/Display Window (Monoxor II) 0019-3055 Front Cover w/Display Window (Nonoxor II) 0019-3078 Front Cover (Monoxor II H) (order with 0019-0437)
Page 19
Single Gas Instrument Service Manual Figure 5-3. Replacement Parts, Back and Front Exploded Views Page 19 Instruction 0019-9166 Bacharach, Inc.
Need help?
Do you have a question about the DIOXOR II and is the answer not in the manual?
Questions and answers