11.4 Checkbox Verification
The Apollo 600 comes supplied with a Checkbox. This can be used to verify that the Apollo
600 test functions are working correctly. When selecting the Checkbox function follow the
onscreen instructions.
The Checkbox function will perform a number of electrical tests. Ensure that
the onscreen instructions are followed and that you do not touch the checkbox
during the active tests.
11.5 PAT Settings
This function allows you to configure the below factors with relation to PAT testing.
Press this button to go to the next page.
Press this button to go to the previous page.
Press this button to save changes and return to the previous screen.
Asset ID
This can be set to present a blank field, repeat the last Asset ID or automatically
increment the Asset ID number for each successive test.
Start
This field is used to set the starting point for the auto increment feature. The
Increment
value entered can be numeric or alpha-numeric ending with a numerical value.
On Test
In the event of a test failure during an automatic sequence, the Apollo 600 can
Failure
be configured to either End Test, which will terminate the test sequence or
present a Fail Menu, which will provide the following list of options:
restart the test
•
skip the test
•
restart the test sequence (from the first test)
•
abort the test sequence (without saving any results
•
end the test sequence (and record a fail result)
•
Note: Restart Test is useful for coping with situations such as forgetting to
connect the earth continuity probe.
End Test Sequence will record a Fail result whereas Abort Test
Sequence does not record a result.
After Test There is a choice of 3 actions to be carried out at the end of a test sequence.
New Test will auto save at the end of a test sequence and then open the Asset
Details window ready for a new test.
Print Label will auto save at the end of a test sequence and then open the Print
Label window. When printing is complete, the tester will proceed to the next test
and open the Asset Details window.
Options Menu will auto save at the end of a test sequence and then open the
Test Details window. Pressing (F4) will proceed to the Test Result Options
window with a list of options including View Results, View Test Sequence,
Need help?
Do you have a question about the Apollo 600 and is the answer not in the manual?