Connect the Pulse Generator Output to the Oscilloscope

U4154A Logic Analyzer Service Guide
Attach Male BNC to Female SMA adapters to Channels 1
1
and 2 on the oscilloscope.
2
Attach one end of an SMA cable to the Male BNC to
Female SMA adapter on Channel 1 of the oscilloscope.
Attach the other end of the SMA cable to the SMA/Flying
3
Lead connector at the Channel 2 OUTPUT of the pulse
generator.
Attach one end of the other SMA cable to the Male BNC
4
to Female SMA adapter on Channel 2 of the oscilloscope.
5
Attach the other end of the SMA cable to the SMA/Flying
Lead connector at the Channel 2 OUTPUT of the pulse
generator.
Testing U4154A Performance
3
37