Adjustment Procedure; Figure 1-8 Reference Value Display - Agilent Technologies U1270 Series Service Manual

Handheld digital multimeters
Table of Contents

Advertisement

Adjustment procedure

N O T E
N O T E
U1270 Series Service Guide
Review the
"Test Considerations"
beginning the adjustment procedures.
1 Turn the rotary switch to the respective test function
position as shown in the adjustment input values table
(Table 1-
3).
2 Unsecure the instrument to enter the adjustment mode.
(See
"Unsecuring the Instrument for Calibration"
page 38).
While in the adjustment mode, press
exit the adjustment mode.
3 The reference value of the calibration item will be shown
on the display.

Figure 1-8 Reference value display

4 Configure each calibration item.
5 Use the arrow keys to select the calibration range.
6 Apply the input signal shown in the Reference value
column of
Table 1-
input reading. There is no bar graph display for
temperature adjustment.
Calibration Procedures
Using the Front Panel for Adjustments
and
"Adjustment considerations"
and
3. The analog bar graph displays the
1
before
on
simultaneously to
47