Agilent Technologies InfiniiVision 3000 X-Series User Manual page 215

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

3 Press the Type softkey to set the measurement threshold to %
Absolute threshold hints
T I P
4 Press the Lower softkey; then, turn the Entry knob to set the lower
5 Press the Middle softkey; then, turn the Entry knob to set the middle
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
Each analog channel and the math waveform can be assigned unique
threshold values.
(percentage of Top and Base value) or to Absolute (absolute value).
Percentage thresholds can be set from 5% to 95%.
The units for absolute threshold for each channel is set in the
channel probe menu.
When the Source is set to Math: f(t), the threshold Type can only be set
to Percent.
Absolute thresholds are dependent on channel scaling, probe attenuation, and probe units.
Always set these values first before setting absolute thresholds.
The minimum and maximum threshold values are limited to on-screen values.
If any of the absolute threshold values are above or below the minimum or maximum waveform
values, the measurement may not be valid.
measurement threshold value.
Increasing the lower value beyond the set middle value will
automatically increase the middle value to be more than the lower
value. The default lower threshold is 10% or 800 mV.
If threshold Type is set to %, the lower threshold value can be set from
5% to 93%.
measurement threshold value.
The middle value is bounded by the values set for lower and upper
thresholds. The default middle threshold is 50% or 1.20 V.
If threshold Type is set to %, the middle threshold value can be set
from 6% to 94%.
14
Measurements
215

Advertisement

Table of Contents
loading

Table of Contents