Both the first and third set of outputs as well as the second and
fourth set of outputs may be set to any of the following functions:
-
THC -- Total hydrocarbon as measured in the first phase
of detection. The output is the percentage of the selected
range.
-
SMP -- An alternate measurement that occurs during the
second phase of detection. The output represents the
percentage of the selected range.
-
WAVE – Analogous to the continuous output of the FID.
-
RID – Indicates the selected range. Low Range is at 1/5
of maximum output. Mid Range is at 1/2 of maximum
output. High Range is 4/5 of maximum output.
Table 3-1: Analog Output Connections
Pin
Function
3
Channel 2, + 4-20 mA, floating
4
Channel 2, – 4-20 mA, floating
5
Channel 1, + 4-20 mA, floating
6
Channel 1, – 4-20 mA, floating
8
Channel 2, 0-1 VDC
23
Channel 2, 0-1 VDC
24
Channel 1, 0-1 VDC
7
Channel 1, negative ground