Cell Failure Checks - Teledyne 3010TB Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzer
The results screen alternates for a time with:
Then the analyzer returns to the initial System screen.
4.5
Cell failure is covered in detail in Part I: Control Units, section 4.4.1.3,
Cell Failure. Cell replacement is covered Part II: Analysis Units chapter 5,
Maintenance.
Cell failure in the 3010T is usually associated with inability to zero the
instrument down to a satisfactorily below 5PPM reading. When this occurs,
the 3010 system alarm trips, and the VFD displays a failure message.
Before replacing the cell:
a. Check your span gas to make sure it is within specifications.
b. Check for leaks downstream from the cell, where oxygen may be
Table 4-1: Self Test Failure Codes
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
Press Any Key
To Continue...

Cell Failure Checks

#.#
CELL FAIL/ ZERO HIGH
leaking into the system.