Humidity Monitor
You can monitor the level of relative humidity over periods of time, print the
results, and store the results in memory. To access the monitor from the
keyboard, enter:
Selections then display as:
RUN HUMIDITY MONITOR A19 TO:
MEMORY
F1
Select
MEMORY
if you want to store the test record. This requires both the RS-
232/Printer module for printing and the 100 Record Memory module for
storage. Select
PRINT
if you want to print the test record. Select
want to store and print the record. Follow these set-up instructions:
1.
Plug the humidity adapter test leads into the
medTester's front panel.
2.
Power on the adapter.
The medTester will prompt you to enter the following information. Entry
information can be any combination of alpha or numeric characters. After
each entry press the
ENT
1.
OP CODE—The operator is the person doing the testing. This
code can consist of a maximum of three characters which can be
entered from the medTester keyboard. An example of an operator
code could be a person's initials or employee number. The
previous operator's code, if any, will appear. If the current
operator has not changed from previous use, press the
2.
LOC—Type in the location of the device, 16 characters
maximum. The previous location code, if any, will appear. If this
occurs and is the desired location, press the
3.
CONTROL #—Enter a control number with a maximum of 16
characters.
TOOLBOX
HUMMON
F2
F5
PRINT
BOTH
F2
F3
key:
Biomedical Test System
Autosequences
NEITHER
F4
BOTH
if you
jacks on
EXT INPUT
ENT
key.
ENT
key.
5
5-21