medTester 5000C
Operators Manual
15-16
For CardioSat 100:
Step Number
(in percent)
1
2
3
4
5
6
7
8
9
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test
Step Number
SPO2
(in percent)
1
96
2
98
3
92
4
82
5
72
6
98
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test
SPO2 Autosequence Number: A70
Autosequence Name: NELLCR
Manufacturer: Nellcor
Settling Time: 25 Beats
Heart Rate Limit:
4 BPM (beats per minute)
SPO2
SPO2 Limit
(in percent)
96
3
98
3
81
4
81
4
81
4
92
3
82
4
72
5
98
3
SPO2 Autosequence Number: A71
Autosequence Name: OHMEDA
Manufacturer: Ohmeda 3700
Settling Time: 25 Beats
Heart Rate Limit:
4 BPM (beats per minute)
SPO2 Limit
Heart Rate
(in percent)
(in BPM)
75
3
230
3
75
3
4
75
75
5
45
3
Heart Rate
Amplitude
(in BPM)
(in percent)
75
100
230
100
38
25
112
100
201
100
75
100
75
100
75
100
30
100
Amplitude
(in percent)
100
100
100
100
100
100