Fluke Biomedical medTester 5000C Operator's Manual page 162

Biomedical tester
Table of Contents

Advertisement

medTester 5000C
Operators Manual
10-24
SECONDARY RATE HIGH LIMIT: Enter the high limit for the
2.
flow rate. 0000.0 turns off high limit checking.
SECONDARY RATE LOW LIMIT: Enter the low limit for the
3.
flow rate. 0000.0 turns off low limit checking.
PRIMARY FLOW RATE: Enter the nominal flow rate, 0000.1 to
4.
0999.0 ml/hr.
PRIMARY RATE HIGH LIMIT: Enter the high limit for the
5.
flow rate. 0000.0 turns off high limit checking.
PRIMARY RATE LOW LIMIT: Enter the low limit for the flow
6.
rate. 0000.0 turns off low limit checking.
TEST END TYPE: Select
7.
MANUAL to end the test by pressing a key.
8.
TIMED to end the test after a specific time:
9.
END TIME: Enter the elapsed time that will end the
10.
test, 00:01 to 99:59 (hours:minutes).
VOLUME to end the test after a specific volume:
11.
END VOLUME: Enter the volume that will end the
12.
test, 0000.1 to 9999.0 ml.
For a PCA Test:
BOLUS VOLUME: Enter the bolus volume, 00.50 to 99.99 ml.
BOLUS VOLUME HIGH LIMIT: Enter the high limit for the
bolus volume. 00.00 turns off high limit checking.
BOLUS VOLUME LOW LIMIT: Enter the low limit for the
bolus volume. 00.00 turns off low limit checking.
LOCKOUT TIME: Enter the lockout time between boluses, 00:01
to 99:99 (minutes:seconds).
END TIME: Enter the elapsed time that will end the test, 00:01
to 99:99 (hours:minutes).
IPT-1
For the IPT-1, there are 2 types of tests available, steady flow and high
accuracy:
FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
1.
ml/hr.
FLOW TEST TYPE: Enter STEADY for steady flow or HI ACC
2.
for high accuracy.
RATE HIGH LIMIT: Enter the high limit for the flow rate.
3.
0000.0 turns off high limit checking.