Using An External Pump; Recommended Pressure Modules - Fluke 718Ex 30G User Manual

Pressure calibrator
Hide thumbs Also See for 718Ex 30G:
Table of Contents

Advertisement

Table 4. Recommended Pressure Modules
External
Pump
Pressure
718Ex
Module
30G/100G/
300G
700P01Ex
X
700P24Ex
X
700P05Ex
X
700P06Ex
X
700P27Ex
X
700P09Ex
X
700PA4Ex
X
700P29Ex
X
Internal Pump
718Ex
718Ex
718Ex
30G
100G
300G
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

Using an External Pump

XW Warning
To avoid damage to the Calibrator and possible
release of pressure, do not connect the internal
sensor to an external pressure source that
exceeds 30 psi for Model 718Ex 30G, 100 psi for
Model 718Ex 100G, or 300 psi for Model 718Ex
300G.
To develop higher pressure or vacuum, use an external
pump. Use a Fluke 700PEx Pressure Module connected to
the pressure module input on the Calibrator. Pressure
modules are listed in Table 4. Make overall connections as
shown in Figure 6.
Refer to setup and operating instructions included with the
pressure module and pump.