Insulation Function, External Sense - Fluke 1587 Calibration Manual

Insulation multimeters
Hide thumbs Also See for 1587:
Table of Contents

Advertisement

3. Turn the rotary switch to INSULATION.
4. Set the voltage range on the UUT; see Table 4.
5. Apply the Calibrator Output to the UUT; see Table 4.
6. Press t on the UUT.
7. Verify that the display reading is within the limits given in Table 4.
UUT
Step
Voltage
Range
1
500 V
2
500 V
3
500 V
4
1000 V
5
1000 V
6
1000 V
7
50 V
8
50 V
9
100 V
10
100 V
11
250 V
12
250 V

Insulation Function, External Sense

The following test verifies that the UUT will sense a voltage > 30 V when present on
the circuit under test.
1. Connect the UUT INSULATION output terminals to the calibrator voltage output
terminals.
2. Turn the rotary switch to the INSULATION function.
3. Apply 35 V, 50 Hz to the UUT.
4. Verify that the UUT displays > 30 V in the primary display.
Table 4. Insulation Resistance Accuracy Test
Calibrator
Calibrator
Resistance
Output
Range
60.0 MΩ
0.500 MΩ
500 MΩ
450 MΩ
500 MΩ
600 MΩ
2.0 GΩ
1.7 GΩ
600 MΩ
600 MΩ
1587 Only
60.0 MΩ
1 MΩ
6.00 MΩ
100 kΩ
50.0 MΩ
45 MΩ
6.00 MΩ
100 kΩ
100 MΩ
95 MΩ
60.0 MΩ
250 kΩ
250 MΩ
225 MΩ
Insulation Multimeters
UUT Display Reading
Limit Low
Limit High
0.0
1.0
438
462
>550
>550
1.2
2.2
586
614
0.5
1.5
0.05
0.15
43.1
46.9
0.05
0.15
87
103
0.0
0.8
217
233
Performance Test
19