Power Saver

The Calibrator automatically turns off after 30 minutes of
inactivity. To reduce this time or disable this feature:
1.
With the Calibrator OFF, press O.
P.S. xx is displayed, where xx is the turn-off time in
minutes. OFF means the power saver is disabled.
h
2.
Press
to decrease or
time.
h
3.
To disable, press
The Calibrator resumes normal operation after 2 seconds.
c
to increase the turn-off
until the display shows OFF.

Zeroing with Absolute Pressure Modules

Zeroing with Absolute Pressure Modules
For zeroing, adjust the Calibrator to read a known
pressure. This can be barometric pressure, if it is
accurately known. An accurate pressure standard can also
apply a pressure within range for any Absolute Pressure
Module. Adjust the Calibrator reading as follows:
Z
1.
Press and hold
.
h
2.
Press
to increase or
Calibrator reading to equal the applied pressure.
Z
3.
Release
to exit the zeroing procedure.
U
Press the
button to convert to any convenient
measurement display unit.