Checking The Diode Test Function - Fluke ProcessMeter 787 Calibration Manual

Hide thumbs Also See for ProcessMeter 787:
Table of Contents

Advertisement

UUT
787
(BLUE)
MIN MAX
% STEP
mV
V
V
OFF
OUTPUT 0-24mA
SOURCE
SIMULATE
+
+
A
mA
30mA
0.44A
FUSED
(1A /30 sec)
FUSED

Checking the Diode Test Function

1. Put the calibrator in Standby (
2. Put the UUT rotary switch in the
3. Press the BLUE button to select diode test (
4. Connect the calibrator to the
Figure 6.
5. Apply 2.0 V dc from the calibrator.
6. The UUT should read between 1.959 V and 2.041 V.
7. Put the calibrator in Standby (
UUT.
8. Put the multimeter in the dc mA (autorange) function.
9. Connect the current terminals of the multimeter to the
on the UUT.
10. The multimeter should read close to 0.2 mA. (There is no tolerance specification for
this current. This test just makes sure that the diode test current source is operating.)
PROCESSMETER
RANGE
HOLD
H
COARSE
FINE
REL
Hz
mA
A
OUTPUT
mA
mA
COM
V
CAT
1000V
Figure 5. Current Measurement Test Connections
STBY
and
COM
STBY
5500A
5500A
CALIBRATOR
SCOPE
NORMAL
AUX
V,
,
A,
-SENSE,
200V PK
RTD
AUX V
STBY
MAX
OPR
EARTH
SCOPE
BOOST
HI
1000V
20V
7
8
9
m
RMS
RMS
MAX
MAX
TRIG
LO
OUT
n
4
5
6
k
1V PK
MAX
20V PK
p
MAX
1
2
3
M
TC
/ +
0
SHIFT
20V PK
MAX
) mode.
position.
.
)
terminals on the UUT as shown in
) mode; then disconnect the calibrator from the
COM
Calibration Manual
Verification Tests
PREV
MENU
dBm
sec
EDIT
V
Hz
SETUP
RESET
FIELD
W
¡F
NEW
CE
A
¡C
REF
MEAS
TRIG
F
TC
OUT
POWER
MULT
DIV
ENTER
x
I
O
LT003F.EPS
and
terminals
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents