Instruction Manual

748297-E
June 2003
Ensure that all gas connections are made as labeled and are leak free. Improper gas connections
could result in explosion or death.
If this Analyzer Module is used with a non-Rosemount Analytical power supply, adding Rosemount
P/N 903341 Current Protector in series with the 24 V positive power line will prevent over-voltage
spiking and resultant fuse blowing when powering up the instrument.
This enclosure shall not be opened unless the area is known to be free of flammable materials or
unless all devices within have been de-energized.
Area classification for the protected enclosure:
Nonclassified.
Pressurization: Type Z
Temperature Identification Number: T4A
Power shall not be restored after enclosure has been opened (or loss of purge) until enclosure has
been purged for a minimum of 6 (six) minutes at the minimum pressure of 689 hPa (10 psig)
Preface
P-4
WARNING
POSSIBLE EXPLOSION HAZARD
CAUTION
OVER-VOLTAGE SPIKING
WARNING
PRESSURIZED ENCLOSURE
HOT INTERNAL PARTS
Rosemount Analytical Inc.
Model NGA2000 HFID
A Division of Emerson Process Management