Summary of Contents for Sierra Innova-Flo 220 Series
Page 1
Sierra Series 220 and 221 Innova-Flo Vortex Flow Meters ™ Instruction Manual Part Number IM-22 05/03 Revision C IMPORTANT: This manual is for use with 220/221 Vortex Products purchased during 1997 through January 2006 CORPORATE HEADQUARTERS 5 Harris Court, Building L Monterey, CA 93940...
Page 2
Unless you have specifically ordered Sierra’s optional O cleaning, this flow meter may not be fit for oxygen service. Sierra Instruments, Inc., is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instru- ments standard mass flow meters for oxygen gas. You are responsible for clean- ing the mass flow meter to the degree required for your oxygen flow application.
Series 220/221 Instruction Manual Table of Contents Table of Contents Chapter 1 Introduction ™ Innova-Flo Vortex Flow Meters............1-1 Using this Manual ................. 1-1 Note and Safety Information ............1-2 Receipt of System Components ............ 1-2 Technical Assistance ..............1-2 How the Innova-Flo Vortex Flow Meter Operates ......
Page 4
Table of Contents Series 220/221 Instruction Manual Chapter 4 Troubleshooting and Repair Hidden Diagnostics Menus..............4-1 Column One Hidden Diagnostics Values........4-2 Column Two Hidden Diagnostics Values ........4-2 Analog Output Calibration ..............4-3 Troubleshooting the Flow Meter ............4-3 Symptom: Output at no Flow ............
Page 5
Always remove main power before disassembling any part of the mass flow meter. Caution! Calibration must be performed by qualified personnel. Sierra Instruments, Inc., strongly rec- ommends that you return your flow meter to the factory for calibration. In order to achieve accurate and repeatable performance, the flow meter must be in- stalled with the specified minimum length of straight pipe upstream and downstream of the flow meter’s sensor head.
Chapter 1 Introduction Chapter 1 Introduction Innova-Flo™ Vortex Flow Meters The Sierra Instruments’ Series 220 In-Line and the Series 221 Insertion Innova-Flo™ Vortex Flow Meters provide a reliable solution for process flow measurement. From a single entry point in the pipeline, Innova-Flo meters offer precise measurements of mass or volumetric flow.
Receipt of System Components When receiving a Sierra flow meter, carefully check the outside packing carton for damage incurred in shipment. If the carton is damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered components are present.
How the Innova-Flo Vortex Flow Meter Operates Figure 1-1. Series 220 In-Line Vortex Flow Meter ™ Sierra Series 220 and 221 Innova-Flo Vortex Flow Meters use a unique velocity sensor head to monitor volumetric flow rate. The built-in flow computer calculates mass flow rate based on a constant value of fluid density stored in the instrument’s memory.
Chapter 1 Introduction Series 220/221 Instruction Manual Vortex Shedding Frequency Von Karman vortices form downstream of a shedder bar into two distinct wakes. The vortices of one wake rotate clockwise while those of the other wake rotate counterclockwise. Vortices generate one at a time, al- ternating from the left side to the right side of the shedder bar.
Page 11
Series 220/221 Instruction Manual Chapter 1 Introduction Flow Velocity Range To ensure trouble-free operation, vortex flow meters must be correctly sized so that the flow velocity range through the meter lies within the measurable velocity range (with acceptable pressure drop) and the linear range.
Chapter 1 Introduction Series 220/221 Instruction Manual ™ As shown below, Innova-Flo Vortex Flow Meters exhibit a constant Strouhal number across a large range of Reynolds numbers, indicating a consistent linear output over a wide range of flows and fluid types. Be- low this linear range, the intelligent electronics in Innova-Flo automati- cally corrects for the variation in the Strouhal number.
Series 220/221 Instruction Manual Chapter 1 Introduction Flow Meter Configurations Innova-Flo™ Vortex Flow Meters are available in two configurations: • Series 220 in-line flow meter (replaces a section of the pipeline) • Series 221 insertion flow meter (requires a “cold” tap or a “hot” tap into an existing pipeline) Both the in-line and insertion configurations are similar in that they both use identical electronics and have similar sensor heads.
Page 14
Chapter 1 Introduction Series 220/221 Instruction Manual The meter includes a local 2 x 16 character LCD display housed within the enclosure. Local operation and reconfiguration is accomplished using six push buttons. For hazardous locations, the six push buttons can be operated through the sealed enclosure using a hand-held magnet, thereby not compromising the integrity of the hazardous location certification.
Series 220/221 Instruction Manual Chapter 2 Installation Chapter 2 Installation Installation Overview Innova-Flo™ meter installations are simple and straightforward. Both the Se- ries 220 In-Line and Series 221 Insertion type flow meter installations are covered in this chapter. After reviewing the installation requirements given below, see page 2-3 for Series 220 installation instructions.
Chapter 2 Installation Series 220/221 Instruction Manual Unobstructed Flow Requirements Select an installation site that will minimize possible distortion in the flow profile. Valves, elbows, control valves and other piping components may cause flow disturbances. Check your specific piping condition against the ex- amples shown below.
Series 220/221 Instruction Manual Chapter 2 Installation Series 220 In-Line Flow Meter Installation Install the Series 220 In-Line Flow Meter between two conventional pipe flanges as shown in Figures 2-3 and 2-4. Table 2-1 provides the recom- mended minimum stud bolt lengths for wafer-style meter body size and dif- ferent flange ratings.
Chapter 2 Installation Series 220/221 Instruction Manual Wafer-Style Flow Meter Installation Install the wafer-style meter between two conventional pipe flanges of the same nominal size as the flow meter. If the process fluid is a liquid, make sure the meter is located where the pipe is always full. This may require lo- cating the meter at a low point in the piping system.
Series 220/221 Instruction Manual Chapter 2 Installation Flange-Style Flow Meter Installation Install the flange-style meter between two conventional pipe flanges of the same nominal size as the flow meter. If the process fluid is a liquid, make sure the meter is located where the pipe is always full. This may require lo- cating the meter at a low point in the piping system.
Chapter 2 Installation Series 220/221 Instruction Manual Series 221 Insertion Flow Meter Installation Prepare the pipeline for installation using either a cold tap or hot tap method described on the following pages. Refer to a standard code for all pipe tap- ping operations.
Series 220/221 Instruction Manual Chapter 2 Installation Cold Tap Guidelines Refer to a standard code for all pipe tapping operations. The following tap- ping instructions are general in nature and intended for guideline purposes only. Caution! 1. Turn off the flow of process gas, liquid or steam. Verify that the line is When using toxic or not pressurized.
Chapter 2 Installation Series 220/221 Instruction Manual Hot Tap Guidelines Refer to a standard code for all pipe tapping operations. The following tap- ping instructions are general in nature and intended for guideline purposes Warning! only. Hot tapping must be performed by a trained 1.
Series 220/221 Instruction Manual Chapter 2 Installation Flow Meter Insertion The sensor head must be properly positioned in the pipe. For this reason, it is important that insertion length calculations are carefully followed. A sensor probe inserted at the wrong depth in the pipe will result in inaccurate readings. Insertion flow meters are applicable to pipes 2 inch and larger.
Chapter 2 Installation Series 220/221 Instruction Manual Installing Flow Meters with a Compression Connection* Use the following formula to determine insertion length for flow meters (NPT and flanged) with a compression process connection. The installation procedure is given on the next page. Insertion Length Formula I = S –...
Series 220/221 Instruction Manual Chapter 2 Installation Insertion Procedure for Meters with a Compression Connection Figure 2-7. Flow Meter with Compression Type Fitting 1. Calculate the required sensor probe insertion length. 2. Fully retract the stem until the sensor head is touching the bottom of the stem housing.
Chapter 2 Installation Series 220/221 Instruction Manual Installing Flow Meters with a Packing Gland Connection* Use the formula below to determine the insertion depth for flow meters (NPT and flanged) equipped with an insertion tool. To install, see the next page for instructions for meters with a permanent insertion tool.
Series 220/221 Instruction Manual Chapter 2 Installation Insertion Procedure for Flow Meters with Permanent Insertion Tool Figure 2-9. Flow Meter with Permanent Insertion Tool 1. Calculate the required sensor probe insertion length (see previous page). Measure from the depth marker arrow down the stanchion and scribe a mark at the calculated insertion depth.
Chapter 2 Installation Series 220/221 Instruction Manual Insertion Procedure for Flow Meters with Removable Insertion Tool Figure 2-10. Flow Meter with Removable Insertion Tool 1. Calculate the required sensor probe insertion length. Measure from the depth marker arrow down the stanchion and scribe a mark at the calcu- lated insertion depth.
Series 220/221 Instruction Manual Chapter 2 Installation 7. Tighten the packing gland nuts to stop leakage around the stem. Do not torque over 20 ft-lbs. 8. Slide the stem clamp back into position. Torque stem clamp bolts to 15 ft-lbs. Replace the stem clamp nuts and torque to 10-15 ft-lbs. 9.
Page 30
Chapter 2 Installation Series 220/221 Instruction Manual Insertion Procedure for Flow Meters with No Insertion Tool (Packing Gland Connection) 1. Calculate the required sensor probe insertion length. 2. Fully retract the stem until the sensor head is touching the bottom of the Warning! The line must be stem housing.
Series 220/221 Instruction Manual Chapter 2 Installation Adjusting Meter Orientation Depending on installation requirements, you may need to adjust the meter orientation. There are two adjustments available. The first rotates the position of the LCD display/keypad and is available on both in-line and insertion me- ters.
Chapter 2 Installation Series 220/221 Instruction Manual Enclosure Adjustment (Series 220 Only) Figure 2-13. Enclosure Viewing Adjustment To avoid damage to the sensor wires, do not rotate the enclosure beyond 180- degrees from the original position. To adjust the enclosure: 1.
Series 220/221 Instruction Manual Chapter 2 Installation Wiring Connections The NEMA 4X enclosure contains an integral wiring compartment with one dual strip terminal block (located in the smaller end of the enclosure). Two Warning! 3/4-inch female NPT conduit entries are available for separate power and To avoid potential electric signal wiring.
Chapter 2 Installation Series 220/221 Instruction Manual 4-20 mA Output Connections The Innova-Flo meter has a single 4-20 mA loop. The 4-20 mA loop current is controlled by the meter electronics. The electronics must be wired in series with the sense resistor or current meter. The current control electronics re- quire 12 volts at the input terminals to operate correctly.
Series 220/221 Instruction Manual Chapter 2 Installation Pulse Output Connections The pulse output is used for a remote counter. When the preset volume or mass (defined in the totalizer settings, see page 3-6) has passed the meter, the output provides a 50 millisecond square pulse. The pulse output requires a separate 5 to 36 VDC power supply.
Chapter 2 Installation Series 220/221 Instruction Manual Remote Electronics Wiring The remote electronics enclosure should be mounted in a convenient, easy to reach location. For hazardous location installations, make sure to observe agency requirements for installation. Allow some slack in the interface cable between the junction box and the remote electronics enclosure.
Press ENTER to continue. (If change is not alllowed, ENTER has no effect.) All outputs are disabled when using the Setup Menus. SIERRA INSTRUMENTS The EXIT key is active within the Setup Menus. When using a Setup Menu, EXIT returns you to the Run Mode.
Chapter 3 Operation Series 220/221 Instruction Manual Start-Up To begin flow meter operation: 1. Verify the flow meter is installed and wired as described in Chapter Note Starting the flow meter or pressing EXIT will 2. Apply power to the meter. At start up, the unit runs a series of self- always display the Run tests that check the program configuration and all flow sensing com- Mode screens.
Series 220/221 Instruction Manual Chapter 3 Operation Using the Setup Menus Run Mode Screens ENTER Volume Password Flow Rate Setup Menus ENTER Mass Flow Rate Output Display Totalizer Units Diagnostics Calibration Password Menu Menu Menu Menu Menu Menu Menu Total 4-20 mA Cycle Time Volume...
Series 220/221 Instruction Manual Chapter 3 Operation Display Menu ENTER Run Mode Password ENTER Display Menu keys to access menus Cycle Time(Sec) If Cycle Time is set to zero, manual advance is required Number of Digits Used to set the number of digits displayed after decimal point Display TC(sec) TC = Display Time Constant, used to smooth display...
Chapter 3 Operation Series 220/221 Instruction Manual Totalizer Menu ENTER Run Mode Password ENTER Totalizer Menu keys to access menus Totaling Example: Inactive Volume Mass Maximum flowrate = 600 gallons per minute (600 gallons per minute = 10 gallons per second) (unit)/Pulse xxxx If unit per pulse is set to 600 gallons per pulse,...
Series 220/221 Instruction Manual Chapter 3 Operation Units Menu ENTER Run Mode Password ENTER Units Menu keys to access menus Volume Flow Unit cm^3 / bbl = barrels = 42 US gallon ft^3 Mass Flow Unit ton = 2000 lb mton = Metric Ton = 1000 kg gram sm^3 = standard cubic meters...
Chapter 3 Operation Series 220/221 Instruction Manual Diagnostics Menu ENTER Run Mode Password ENTER Diagnositcs Menu keys to access menus Sim Vor Freq Simulate Vortex Frequency (Hz) Highest Velocity Highest Recorded Velocity (ft/sec) Use the Diagnostics Menu to simulate flow and review the highest recorded veloc- ity in ft/sec.
Series 220/221 Instruction Manual Chapter 3 Operation Calibration Menu ENTER Run Mode Password ENTER Calibration Menu keys to access menus Meter Size Series 220 - meter size or Pipe ID Series 221 - pipe internal diameter (inches) Meter calibration constant Meter Factor Series 220 - pulses/ft xxxx...
Chapter 3 Operation Series 220/221 Instruction Manual Password Menu ENTER Run Mode Password ENTER Password Menu keys to access menus Set Password 1234 Use the Password Menu to set or change the system password. The factory-set password is 1234. 3-10 IM-22-C...
Series 220/221 Instruction Manual Chapter 4 Troubleshooting & Repair Chapter 4 Troubleshooting and Repair 4-20, Zero xxxx Vper 4-20, FScale Vcyc xxxx Vor Freq Direct? Disable Recorct Gain Control Filter Control Not Present on Pre-Filt 220 models Water? Total in EEprom Config Code Factory...
Chapter 4 Troubleshooting & Repair Series 220/221 Instruction Manual Column One Hidden Diagnostics Values • f = vortex shedding frequency (Hz). • fi = adaptive filter – should be approximately 25% higher than the vortex shedding frequency, this is a low-pass filter. •...
Series 220/221 Instruction Manual Chapter 4 Troubleshooting & Repair • Pre-Filt Water? = Pre-filter for liquid applications. YES for all liq- uid applications and NO for all gas or steam applications. Always set to NO for ½” and ¾” meters due to the high vortex shedding fre- quency.
Chapter 4 Troubleshooting & Repair Series 220/221 Instruction Manual 2. Mechanical installation may be incorrect. Verify the straight run is adequate as described in Chapter 2. For in-line meters, make sure the meter is not installed backwards and there are no gaskets protrud- ing into the flow stream.
Series 220/221 Instruction Manual Chapter 4 Troubleshooting & Repair analog board. The meter should read electrical noise, 60 Hz for ex- ample. If all readings are correct, re-install vortex sensor wires. 4. Verify all meter configuration and troubleshooting steps previously described.
Chapter 4 Troubleshooting & Repair Series 220/221 Instruction Manual Returning Equipment to the Factory Before returning any Innova-Flo meter to the factory, you must re- quest a Return Material Authorization (RMA) number. To obtain an RMA number and the correct shipping address, contact Customer Ser- vice at: (800) 866-0200 or (831) 373-0200 in the USA, or +31(0)20-6145810 in Europe.
Series 220/221 Instruction Manual Appendix A Specifications Appendix A Product Specifications Accuracy Process 220 Series In-Line Meters 221 Series Insertion Meters Variables Liquids Gas & Steam Liquids Gas & Steam Volumetric ±0.7% of rate ±1% of rate ±1.2% of ±1.5% of rate Flow Rate over a over a...
Page 54
Appendix A Specifications Series 220/221 Instruction Manual Water Minimum and Maximum Flow Rates ½-inch ¾-inch 1-inch 1.5-inch 2-inch 3-inch 4-inch 6-inch 8-inch 1076 2437 4270 Linear Range Smart electronics corrects for lower flow down to a Reynolds num- ber of 5,000. The Reynolds number is calculated using the fluid density and viscosity entered into the memory.
Page 55
Series 220/221 Instruction Manual Appendix A Specifications Power Requirements Loop powered, 12 to 36 VDC. Output Signals Analog: field rangeable linear, 4-20 mA output signal, 1000 ohms maximum loop resistance, selected by user for volumetric flow rate or mass flow rate. Pulse: field rangeable volume/pulse output for totalization is a 50- millisecond duration pulse operating a solid-state relay capable of switching 40 VDC, 40 mA maximum.
Page 56
Appendix A Specifications Series 220/221 Instruction Manual IM-22-C...
Series 220/221 Instruction Manual Appendix B Glossary Appendix B Glossary A B C D Cross sectional area. acfm Actual cubic feet per minute (volumetric flow rate). ASME American Society of Mechanical Engineers. Bluff Body A non-streamlined body placed into a flow stream to create vortices.
Page 58
Appendix B Glossary Series 220/221 Instruction Manual I J K L In-Line Flow Meter A flow meter which includes a short section of piping which is put in-line with the user’s piping. Insertion Flow Meter A flow meter which is inserted into a hole in the user’s pipeline.
Page 59
Series 220/221 Instruction Manual Appendix B Glossary Q R S T Flow rate, usually volumetric. Rangeability Highest measurable flow rate divided by the lowest measurable flow rate. Reynolds Number A dimensionless number equal to the density of a fluid or Re times the velocity of the fluid times the diameter of the fluid channel, divided by the fluid viscosity (i.e., Re = ρVD/µ).
Page 60
Appendix B Glossary Series 220/221 Instruction Manual IM-22-C...
Need help?
Do you have a question about the Innova-Flo 220 Series and is the answer not in the manual?
Questions and answers