Page 1
Integrated Instruction Manual WP 600 Series Meters pH 600/ 610/ 620, COND 600/610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Rev. 1 Oct 07 Technology Made Easy ... Part of Thermo Fisher Scientific...
Page 3
*IMPORTANT – PLEASE READ BEFORE YOU CONTINUE* PLEASE USE THIS MANUAL IN THIS ORDER: Step 1: Use the first half of this manual “General Guide” to set up and calibrate your instrument. [Including the temperature parameter guide] (Pages 1 - 35) Step 2: Then go to the specific parameter guides (eg.
Table of Contents GENERAL GUIDE..................1 1. Overview – For All Meters ..................3 1.1 About the Meters..........................3 1.1.1 Display Overview........................3 1.1.2 Key Functions.......................... 3 1.2 Inserting Batteries ..........................4 1.2.1 Inserting batteries for the first time..................4 1.2.2 Changing batteries subsequently................... 4 1.3 Attaching the Belt ..........................
Page 6
2.1.1 About pH Calibration ......................36 2.2 pH buffer group for calibration and auto-recognition ..............37 2.2.1 pH Calibration points ......................37 2.3 pH Calibration with a Standard Buffer..................... 37 2.4 pH Calibration with a User-defined Buffer..................39 2.5 Calibration Report ........................... 40 2.6 Average Slope Indicator of pH Probe .....................
WP 600 Series Meters. Please do not skip this section. Model Parameters pH 600 Temperature (with ATC), pH (-2.00 to 20.00 measuring range). pH 610 Temperature (with ATC), pH (expandable resolution to 0.001). pH 620 Temperature (with ATC), pH (expandable resolution to 0.001) and Ion. COND 600 Temperature (with ATC), Conductivity, TDS.
1. Overview – For All Meters 1.1 About the Meters SPECIAL FEATURES • Displays and measures up to 4 parameters simultaneously • Automatic temperature compensation • Built in memory backup to save calibration and 500 sets of measured data • Data logging feature date-and-time stamp to meet Good Laboratory Practice (GLP) •...
1.2 Inserting Batteries NOTE:Please ensure that the gasket is in place otherwise the instrument will not be waterproof. Power up your meter using either: 1. Four ‘AA’ size 1.5 V alkaline batteries (supplied) or, 2. 9V DC power adapter (Optional in some models). 1.2.1 Inserting batteries for the first time 1.
1.3 Attaching the Belt To attach the safety belt: 1. Use a Phillips screw driver to remove the four screws holding the battery cover. 2. Insert the safety belt through the two slots as indicated below. 3. Screw the battery cover back on. Note the ▲UP symbol marked on the cover. 4.
1.4 Connecting Peripherals 1.4.1 Probes (pH/Conductivity/DO) Attach the probes with correct type of connectors as indicated. pH Probe (BNC) DO Probe with built in temperature sensor (6-pin Connector) Conductivity Probe with built in temperature sensor (8-pin Connector) Use this socket for the standalone temperature probe in the pH only...
1.4.2 Protective Rubber Boot The rubber boot protects the meter when used in the field. For bench top applications, lift up the stand at the back of the rubber boot. PLEASE SET UP THE SYSTEM BEFORE YOU BEGIN USING THE METER. USE THE FOLLOWING INSTRUCTIONS: 2.
Page 16
Parameter Description Factory Default STABLE ENABLE - The meter displays ‘Stable’ indicator in the ENABLE Indicator measurement screen as per the ‘STABLE CRITERIA’ defined below. DISABLE – ‘Stable’ indicator does not appear. Stability FAST SLOW – The reading is stabilized slowly and exhibits Criteria good repeatability MEDIUM –...
Page 17
System Settings – Page 2 Figure 1 : System Settings - Page 2 This page allows you to set the date & time of the meter. Parameter Description Factory Default Year Sets the current year 2006 Month Sets the current month Date Sets the current date Hour...
Page 18
Parameter Description Factory Default Auto OFF ENABLE – Turns the meter off automatically if no key is ENABLE pressed for the time period specified in ‘ON TIME’ below. However, this will happen only if you are using the battery, NOT when the meter is plugged into an AC power source or when it is printing data.
Page 19
This allows you to set wireless serial data communication related parameters. Parameter Description Factory Default Print Mode IrDA – Sets serial data communication protocol to IrDA IrDA LED – Sets serial data communication protocol to RS232C MEM- Logs data to meter’s memory. Data Format CyberComm CyberComm –...
Page 20
However, if the user forgets his password, he can contact the nearest distributor or Eutech Instruments/Oakton Instruments to request for meter password. This would be unique to each instrument and would be tied to the serial number of the unit.
Page 21
System Settings – Page 6 This allows you to clear the memory and reset the meter to factory defaults. Parameter Description Factory Default Clear logged YES – Select this to clear all the stored data from the Data Memory meter’s memory NO –...
OVERVIEW FOR SETUP MODE 1. Setup Mode – For All Meters About Setup Mode The setup mode lets you configure various parameters & settings of the meter. You can choose to password-protect your settings, so that other users who may use the meter will not be able to change the settings.
Function Keys available in setup key function screen: To select individual setup ENTER To select or confirm the selection. NEXT-P To navigate to next page. NEXT To go to next parameter without saving the changed parameter. To go back to measurement mode. 4.
Figure 5: Login password screen Note: You can enter ‘00000’ (read-only password) if you wish to view the setup parameters. You are not allowed to modify any parameter when you enter ‘read-only password’. 4. Press up & down arrow keys to enter the first digit of the password and then press NEXT (F3) key to move to the next digit.
Page 25
• Press ENTER key to save the change, or • Press NEXT (F3) key to go to the next parameter without saving the changed parameter. 4. Press NEXT-P (F2) or PRE-P (F1) to navigate to next or previous page. 5. Press ESC (F4) to exit from setup mode. Function Keys available in setup sub-group screens: PRE-P (F1) Goes to the previous page of the same sub-group...
ENTER Selects the current sub-group ESC (F4) Goes to measurement mode from where you entered setup (Not functional) OVERVIEW FOR CALIBRATION MODE 2. Calibration Mode - For All Meters 2.1 About Calibration The 600 series meters are factory calibrated and allows you to measure pH/mV/ion/conductivity/resistivity/TDS/salinity/DO(%)/DO(mg/L) respective to the model(s) you have purchased.
Page 27
For pH Connect the pH probe to the BNC connector of the meter. Be sure to remove the protective electrode storage bottle or rubber cap of the electrode before calibration or measurement. If the electrode has been stored dry, wet the electrode in clean water for 10 minutes before calibrating or taking readings to saturate the pH electrode surface and minimize drift.
For DO(%) and DO (mg/L) Calibration Before starting calibration, make sure you are in the correct measurement mode and in the correct calibration sequence. The temperature and the % Saturation calibration must be done first before attempting to do the mg/L (ppm) Concentration calibration.
Figure 7 : Login password screen Note: You can enter ‘00000’ (read-only password) if you wish to view the calibration report of the last calibration. You are not allowed to perform calibration when you enter ‘read-only password’. 3. Press up &...
pH measurement mode mV measurement mode Ion measurement mode Conductivity measurement mode TDS measurement mode Salinity measurement mode Resistivity measurement mode % - DO percentage saturation measurement mode mg/L(ppm) - DO concentration mode Only the PCD 650 model is equipped with all of them. The meter automatically goes to the mode that was used before it was turned off the last time.
Function Keys available in measurement screen (2 Group): SETP (F1) Goes to setup mode MEM (F2) Shows stored data in the memory STOR (F3) Stores the currently displayed reading in the memory ESC (F4) Shows 1 Group of functions Function Keys available in measurement screen (3 Group): REPO(F2) Shows corresponding calibration report (based on selected measurement...
3.2.4 Holding a reading In some situations, you may want to freeze (hold) the measured reading in the LCD for a delayed observation. You can hold a reading in two different ways: Manual Hold – Allows you to hold the reading by pressing HOLD (F1) key at any time you want.
3.4 Alarm set point (For pH/conductivity/DO) meter display alarm when pH/conductivity/salinity/resistivity/TDS/DO(%)/DO(mg/L) reading goes higher or lower than predefined set points. Refer to ‘Setup’ pages of respective modes to set alarm points for pH, conductivity and DO. Figure 9 : Hi alarm 3.5 Calibration Due (CAL-DUE) Indicator (For pH/conductivity/DO) You can set a reminder to be displayed in the LCD when the next calibration is due.
TEMPERATURE PARAMETER GUIDE i. Temperature Setup Mode ii. Temperature Calibration – ATC and MTC Mode iii. Temperature Measurement Mode FOR ALL MODELS : pH 600/ 610/ 620 COND 600/610 DO 600 PC 650 PD 650 CD 650 PCD 650...
1. Temperature Setup Temperature setup sub-group allows you to configure temperature measurement & calibration related settings of the meter. Temperature Setting Page Figure 10 : Temperature Settings Page Parameter Description Factory Default Unit ºC Sets the unit of measurement for temperature. Available units: ºC and ºF Mode Sets the temperature compensation mode.
Calibrate the probes only if you suspect temperature errors may have occurred over a long period of time or if you have a replacement probe. This procedure offers offset adjustment of the probe to ensure more accurate temperature measurement. Use a thermometer you are certain is accurate to measure the temperature of your sample.
6. Press ENTER key to confirm temperature value. 2.3 Temperature Calibration for MTC mode Make sure you have selected ‘MTC’ and required unit of measurement (ºC or ºF) in Temperature settings. Switch on the meter. Make sure the meter is in measurement mode.
1. pH Setup pH setup sub-group allows you to configure pH measurement & calibration related settings of the meter. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages. pH Settings – Page 1 Figure 11: pH Settings - Page 1 This page allows you to set pH measurement &...
pH Settings – Page 2 Figure 12: pH Settings - Page 2 Parameter Description Factory Default pH Calibration Specify number of days for the pH calibration alarm. 5 Days The meter shows CAL DUE indicator after calibration due days are passed from the last calibration date. Available range: 0 to 30 0.00 pH Resolution...
The meter is capable of calibrating up to 6 points, depending on the standard buffer selected. When completely re-calibrating the meter, the first point of calibration should be one of the following, depending on your choice of standard buffer group. First Point of Calibration Other Available Calibration Points Buffer...
Page 46
Figure 13: pH Calibration Screens 4. Rinse the electrode in clean water and prepare electrode for calibration. Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data. Press CLR-C (F2) key to clear previous calibration.
2.4 pH Calibration with a User-defined Buffer If you selected ‘USER’ (Custom) buffer in the pH Setup, the following screen is shown when you enter calibration mode. You need to prepare at least 2 custom buffers of known pH values. Note: Custom buffer solution values should be at least 1 pH unit apart from each other.
Function Keys available in pH Calibration mode: TEMP (F1) Goes to temperature calibration CLR-C (F2) Clears previous calibration (if any) after ENTER key is pressed for the confirmation screen NEXT (F3) Shows calibration report ESC (F4) Exits from calibration and goes back to pH measurement mode ENTER Confirms the calibration Increases/decreases pH buffer reading...
The slope is 88 for the range enveloped from pH 4.01 to pH 7.00. The slope is 94 for the range enveloped from pH 7.00 to pH 12.45 where pH 10.01 calibration point has been skipped. Function Keys available in pH calibration report screen: PAGE (F1) Goes to the second page of the calibration report BACK (F1)
3. pH Measurement Mode 3.1 Resolution of pH Reading By default, the resolution is set to 0.00. You can change the resolution in the setup mode. pH resolution can be adjusted from 0.0 to 0.000 pH. 3.2 Indicators in pH measurement screen In pH measurement mode, the meter displays pH and temperature reading.
1. mV Setup Mode There is no setup mode for mV parameter. 2. mV Calibration Mode Use a standard ORP solution of known value for calibration. You need to calibrate only 1-point. 1. Switch the meter on. Make sure the meter is in mV measurement mode. 2.
7. Press ESC (F4) to go to measurement mode. The meter shows the relative mV reading of the solution in measurement mode, if offset is not zero. Note: The function of the keys available on the mV calibration/ report screen is same as previously described in pH calibration with a user- defined buffer section.
Page 55
Item Description Measurement mode indicator Appears when the reading is stable Appears when the reading is on hold mV reading Units of measurement Temperature reading & units Temperature compensation mode Note: If offset is not zero, the meter displays relative mV reading of the solution in measurement mode.
1. Ion Setup Ion setup sub-group allows you to configure Ion measurement & calibration related settings of the meter. Ion Settings Page Figure 17: Ion Settings Page Parameter Description Factory Default Ion unit Sets the unit of measurement for Ion. Available units: ppm, molar and mg/L Press (Up) or...
Page 60
Use standard solutions of 0.001, 0.01, 0.1, 1, 10, 100, 1000 & 10000 ppm for calibration. You need to calibrate minimum of 2-points. See ‘Error! Reference source not found.’ in previous section. 1. Switch on the meter. Make sure the meter is in Ion measurement mode. 2.
7. The upper display shows the next calibration point. 8. Rinse the ISE with clean water. Place it in the next calibration solution and follow the steps 4 through 6 to calibrate other points. Note: The meter accepts the second and subsequent points only if the mV readings are within the calibration window (15mV/decade to 90mV/decade).
3. Ion Measurement Mode In Ion measurement mode, the meter displays Ion concentration (in ppm, molar or mg/L) and mV reading. Depending on application, you can connect a suitable Ion Selective Electrode (ISE) to the meter. 1.1 Changing unit of measurement By default the measuring unit is ppm.
1. Conductivity Setup Conductivity setup screen present many options to control the operating parameters, which can be controlled and set from the conductivity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
Page 66
Note: 1. If pure water coefficient is set to ‘ENABLE’, the meter will apply pure water compensation automatically below 2 switch back to linear μS and compensation for above 2 μS. 2. If pure water coefficient is set to ‘DISABLE’, Lnr Temp Coefficient applies throughout the measurement.
2. Conductivity Calibration Mode 2.1 Conductivity Calibration mode This option allows you to select automatic or manual calibration standard. In the automatic calibration mode, the meter automatically detects and verifies the appropriate known calibration standards solutions before accepting these particular calibration standards as one of its calibration values in a specific measurement range.
2.5 Pure Water Coefficient Pure water coefficient will be calculated and applied automatically for ultra pure water measurement if you select ‘ENABLE’ at appropriate set up pages. 2.6 To begin Calibration Make sure you have selected a standard solution with which you intend to perform conductivity calibration.
2.7 Manual Calibration Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data. Press CLR-C (F2) key to clear previous calibration. The meter shows you confirmation screen. Press ENTER key to confirm deleting previous calibration. Figure 25 : Conductivity calibration Screen 7.
Page 70
The following screen is displayed after cell adjustment screen, if you have selected ‘Auto’ in Calibration mode and ‘MULTI’ in Calibration Point: Figure 26 : Conductivity calibration Screen-Auto 1. Rinse the electrode in de-ionized water and prepare electrode for calibration. Note: If you wish to completely re-calibrate the meter, you need to clear previous calibration data.
ESC (F4) Exits from calibration and goes back to conductivity measurement mode ENTER Confirms the calibration Not functional 6. Press ESC (F4) to go to measurement mode. Note: If the user has selected ‘SINGLE’ point calibration and ‘AUTO’ calibration mode, the meter will go to measurement mode after step 4. 2.9 Calibration Report Calibration report gives you detailed information on calibration.
3. Conductivity Measurement Mode In conductivity measurement mode, the meter displays conductivity and temperature readings. The LCD shows related information for the conductivity measurement such as temperature compensation mode, stable indicator, probe condition, calibration points, response time of the conductivity probe and conductivity alarm conditions. You can customize or enable/disable some of these indicators in the Setup mode.
1. TDS Setup The TDS setup screen presents many options to control the operating parameters, which can be controlled and set from the TDS setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
Page 76
TDS Settings – Page 2 Figure 30 : TDS Settings - Page 2 This page allows you to set alarm limits for the TDS measurement mode: Parameter Description Factory Default Alarm Set Set alarm limits for the TDS measurement mode: DISABLE Points ENABLE –...
2. TDS Calibration Mode You can commence calibration in the TDS mode by using the conductivity calibration method and after setting the correct TDS factor. For the rest of the calibration process, follow steps as mentioned in the section on manual calibration. Use a standard solution of known value for calibration.
Note: The function of the keys available on the TDS cell constant/calibration/ report screen is the same as previously described in the conductivity calibration section. 8. Rinse the probe with de-ionized water. Place it in the next solution and follow the steps 2 to 7 to calibrate other points, if ‘MULTI’...
Page 79
Item Description Measurement mode indicator Appears when the reading is stable Appears when the reading is on hold TDS reading Units of measurement Temperature reading & units Temperature compensation mode TDS HI & LO Alarm limits TDS Alarm indicator Response time Calibration Due indicator Number of calibration points and method of calibration...
1. Salinity Setup Salinity setup screen presents many options to control the operating parameters, which can be controlled and set from the salinity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
This page allows you to set alarm limits for the salinity measurement mode: Parameter Description Factory Default Alarm Set DISABLE Set alarm limits for the salinity measurement mode: Points ENABLE – The measurement screen shows HI or LO alarm indicators when the meter reading is above ‘HI alarm value’...
4. The upper display of the cell constant adjustment screen shows the cell constant of your cell and lower display shows the measured value of the solution with adjusted cell constant. Press & keys to adjust the cell constant for your custom solution in upper display. 5.
3. Salinity Measurement Mode In Salinity measurement mode, the meter displays salinity and temperature reading. 3.1 Indicators in salinity measurement mode Note: Lower display shows ‘2Cell’ or ‘4Cell’ when a 2-cell or 4-Cell conductivity probe is connected Figure 37 : Salinity measurement screen Item Description Measurement mode indicator...
1. Resistivity Setup The resistivity setup screen presents many options to control the operating parameters, which can be controlled and set from the resistivity setup screen. The settings are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through these pages.
This page allows you to select alarm limits for resistivity measurement mode. Parameter Description Factory Default Alarm Set Set alarm limits for the resistivity measurement mode: DISABLE Points ENABLE – The measurement screen shows HI or LO alarm indicators when the meter reading is above ‘HI alarm value’...
Note: If the meter is password protected, you will be prompted to enter a password. Figure 40 : Resistivity Cell constant & resistivity calibration 4. The upper display of the resistivity cell constant screen shows the cell constant of your cell & lower display shows the measured value of the solution with adjusted cell constant.
2.1.1 To View resistivity Report: 1. From resistivity measurement mode, press left or right arrow key to navigate to other available functions until you see REPO function in the 2. Press REPO (F2) key. The calibration report is shown in the display 3.
Page 93
Item Description Measurement mode indicator Appears when the reading is stable Appears when the reading is on hold Resistivity reading Units of measurement Temperature reading & units Temperature compensation mode Resistivity HI & LO Alarm limits Resistivity Alarm indicator Response time Calibration Due indicator Number of calibration points and method of calibration...
PART A - % Saturation Mode 1. O2 % - DO Saturation Setup The DO saturation (%) setup screen presents many options to control the operating parameters, which can be controlled and set from the DO setup screen. The settings are displayed in 2 pages.
Page 98
NOTE: During measurement, the meter will automatically measure and compensate for barometric pressure. However, if the user feels that barometric pressure is inaccurate and needs to be calibrated, it can be adjusted in the DO setup screen. DO (%) – Page 2 Figure 44: DO (%) - Page 2 This page allows you to set alarm limits for the DO saturation measurement mode.
2. DO Calibration in % Saturation Mode (with ATC) 2.1 About DO(%) and DO (mg/L) Calibration The calibration of % Saturation of DO will linearly affect the measurement for DO in mg/L. The amount of oxygen dissolved in water will depend on its temperature, atmospheric pressure and its salinity.
Page 100
3. The meter shows the ‘Dissolved O Calibration-Rinse Electrode’ screen for a few seconds to prompt user to rinse electrode with de-ionized water before calibration. Figure 45: Saturation Calibration Screens 4. Rinse the probe well with de-ionized water. For best results, blot the end of the probe to dry it.
Figure 46 : Saturation measurement Screen Function Keys available in saturation calibration mode: TEMP (F1) Goes to temperature calibration CLR-C (F2) Clears previous calibration (if any) after ENTER key is pressed in confirmation screen NEXT (F3) Shows calibration report ESC (F4) Exits from calibration and goes back to saturation measurement mode ENTER Confirms the calibration...
1. Press MODE key to select % Saturation mode. 2. Immerse the DO probe in 0% solution. Stir gently to create a homogenous solution. 3. Press CAL key to calibrate the meter. The meter shows the ‘Dissolved O Calibration-Rinse Electrode’ screen for a few seconds to prompt user to rinse electrode with de-ionized water before calibration.
Figure 48: DO Report in % saturation mode Function Keys available in saturation report screen: PRIN (F2) Transfers calibration report to Computer through IrDA NEXT (F3) Goes to saturation measurement mode ESC (F4) Goes to saturation measurement mode ENTER (Not functional) (Not functional) 2.4 % Saturation Offset Adjustment This option lets you offset the meter’s value when cross referenced with...
3. Percentage Saturation (%) Measurement Mode In percentage saturation measurement mode, the meter displays % saturation and temperature reading. The LCD shows related information for the saturation measurement such as temperature compensation mode, stable indicator, barometric pressure, calibration points, probe condition, response time of the DO probe and alarm conditions.
PART B – Concentration (mg/L) (ppm) Mode 1. O2 mg/L (ppm) – DO Concentration Setup DO Concentration (mg/L) setup screen present many options to control the operating parameters, which can be controlled and set from the DO setup screen. The settings are displayed in 2 pages.
Page 106
NOTE: Auto salinity compensation will be applicable only in multi measurement mode and only if one of the measurement selected is conductivity related. Otherwise, a manually set salinity value will be applied. DO (mg/L) – Page 2 Figure 51: DO (mg/L) – Page 2 This page allows you to set alarm limits for the DO concentration mode.
2. DO Calibration in mg/L or ppm Concentration Mode The amount of oxygen dissolved in a liquid will depend on its temperature, atmospheric pressure and salinity. It is therefore very important to set temperature and salinity correctly before attempting to do a calibration. The meter is capable of measuring barometric pressure with its built-in pressure sensor.
7. Press ENTER key to confirm the entered value. 8. Press ESC (F4) to go to measurement mode. The meter shows the concentration reading of the solution in measurement mode. Figure 53 : Concentration calibration Note: The function of the keys available on the concentration calibration/ report screen is same as previously described in saturation calibration.
2.2 Set Salinity This option allows you to set the salinity (in ppt) for your solution, as the dissolved salts alter the relationship between the partial pressure of oxygen and the oxygen concentration in water. By setting the salinity, the meter will compensate for the impact of salinity on the partial pressure/oxygen concentration relationship of your sample.
Page 110
Appears when the reading is stable Appears when the reading is on hold Concentration reading Units of measurement Temperature reading & units Temperature compensation mode HI & LO Alarm limits Alarm indicator Figure 54 : ppt indicator for input of salinity value DO calibration point Concentration measurement screen -END OF D.O.
1. TRANSFERRING AND PRINTING DATA Please refer to this section after completing calibrations and taking measurements. 1.1 CyberComm 600 Data Acquisition Software Your meter is shipped with a companion software application called CyberComm 600 Data Acquisition Software (DAS). For details on installation, connection and usage of the software, please read the manual that comes with the software CD.
1.2.2 Storing a current measurement (In IrDA and LED print mode: Applicable to all modes) 1. Make sure you are in the measurement mode. 2. Press left or right arrow key to navigate to other available functions until you see STOR function in the LCD. 3.
Page 113
4. Press PRIN (F2). The screen appears for you to select printing options. 5. Press up or down arrow key to select either ‘all (memory) locations’ or the ‘current memory location’, and then press ENTER key. 6. If you have selected ALL LOCATIONS, then you can specify the time interval in between each transfer.
Transfer Stored data to CyberComm 1.2.5 Transferring stored data to a PC not equipped with infrared receiver. If your PC doesn’t come with a built-in infrared receiver, you can use an RS 232C Adapter (sold separately) that connects directly to your PC using a specially-designed 9-pin plug.
1.2.6 Transferring stored data to a PC using an USB/ irDA Dongle. You can purchase an USB/ irDA dongle 01X447601 (see accessories page at the back of this manual) and transfer data from your meter using the USB port. Please note: This accessory is specially engineered to match your meter’s software to the application in your laptop or PC.
-10.0 ºC to 60.0 ºC (14.0 ºF to 140.0 ºF) Resolution 0.1 ºC/ 0.1 ºF Relative Accuracy ± 0.5 ºC / ± 0.9 ºF Model pH 600 pH 610 pH 620, PD 650,PCD 650 Range -2.00 to 20.0 pH -2.000 to 20.000 pH Resolution 0.1/ 0.01 pH 0.1/0.01/0.001 pH...
Page 117
Conductivity Sub range Range Measuring ranges Resolution Accuracy Standards 0.050uS to 2.000uS 0.01uS* 1% of FS 2.000uS to 9.990 uS 0.01uS* 1% of FS 10.00uS to 99.99uS 0.01uS 1% of FS 84.00uS 100.0uS to 300.0uS 0.1uS 1% of FS 300.0uS to 999.9uS 0.1uS 1% of FS 1.413mS...
Page 118
* Display resolution is 0.001 but actual resolution shows 0.01. Resistivity Sub range Range Measuring ranges Resolution Accuracy 10.00MΩ to 20.00MΩ 0.01MΩ 1% of FS 1.000MΩ to 9.999MΩ 0.001MΩ 1% of FS 500.0KΩ to 999.9KΩ 0.1KΩ 1% of FS 100.0KΩ to 500.0KΩ 0.1Ω...
Page 119
Barometric Pressure Resolution 1 mmHg Barometric Pressure Accuracy ± 1% Salinity Correction 0 to 50 ppt Cal Due Alarm Yes (max -30 days) Set Point Alarm Input DO 6 Pin Round Display Display type Dot matrix LCD with backlighting Screen resolution 110 x 128 Viewing area 68 x 74 mm...
1. APPENDIX 1.1 Conductivity theory Conductance is a quantity associated with the ability of primarily aqueous solutions to carry an electrical current, I, between two metallic electrodes when a voltage E is connected to them. Though water itself is a rather poor conductor of electricity, the presence of ions in the water increases its conductance considerably, the current being carried by the migration of the dissolved ions.
Page 123
The salinity scale which ranges from 0 to 80 ppt is a measure of all salts, not just sodium chloride. This scale was originally devised for seawater, and is based on seawater at 15 degrees Centigrade has a conductivity equivalent to that of a potassium chloride solution of a known concentration.
1.2 Calculating TDS Conversion Factor You can calibrate your meter using TDS calibration standard solutions. The calibration standard only needs to give the TDS value at a standard temperature such as 25 °C. To determine the conductivity-to-TDS conversion factor use the following formula: Factor = Actual TDS ÷...
1.4 Dissolved Oxygen Probe 1.4.1 Dissolved Oxygen Principle The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measurements immediately – without the typical 15 minute wait of other dissolved oxygen probes.
1.4.3 Membrane Housing Replacement Replacement of the membrane cap housing/ membrane is required only when you cannot calibrate the probe, or if the membrane is damaged. Typical membrane damages are punctures or wrinkles caused during measurements or cleaning. Your new DO probe comes with replacement membrane housing. To order more replacement membrane housing, see the “Accessories”...
2. Hold the probe under hot running water and brush away the white oxide on the cylindrical anode with a stiff plastic brush – do not use metal cleaning material. 3. If the cathode has any deposits, remove them with a light scouring powder. Do not polish the cathode.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Page 130
For more information on Eutech Instruments’/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments Eutech Instruments Pte Ltd. Distributed by: P.O Box 5136, Blk 55, Ayer Rajah Crescent, Vernon Hills, IL 60061, USA...
Need help?
Do you have a question about the pH 610 and is the answer not in the manual?
Questions and answers