General Specifications; Voltage; Dimensions; Display - Fluke 8808A User Manual

Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

8808A
Users Manual

General Specifications

Voltage

100V Setting ...................................................... 90 V to 110 V
120V Setting ..................................................... 108 V to 132 V
220V Setting ..................................................... 198 V to 242 V
240V Setting ..................................................... 216 V to 264 V
Frequency.......................................................... 47 Hz to 440 Hz
Power Consumption ......................................... 15 VA peak (10 W average)

Dimensions

Height................................................................. 88 mm (3.46 in)
Width .................................................................. 217 mm (8.56 in)
Depth ................................................................. 297 mm (11.7 in)
Weight ................................................................ 2.1 kg (4.6 Ib)

Display

Vacuum Fluorescent Display, segment

Environment

Temperature
Operating ........................................................ 0 C to 50 C
Storage ........................................................... -40 C to 70 C
Warm Up......................................................... ½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating ........................................................ <90 % (0 C to 28 C)
Storage .......................................................... -40 C to 70 C <95 %
Altitude
Operating ....................................................... 2,000 Meters
Storage............................................................... 12,000 Meters
Vibration ............................................................ Complies with MIL-PRF-28800F Class 3

Safety

Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V

EMC

Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000

Triggering

Trigger Delay .................................................... 400 ms
External Trigger Delay ..................................... <2 ms
External Trigger Jitter ..................................... <1 ms
Trigger Input ..................................................... TTL Levels
Trigger Output ................................................... 5 V max

Math Functions

Min/max, relative, hold, compare and dB functions

Electrical

Input Protection ............................................... 1000 V all ranges
1-8
<75 % (28 C to 40 C)
<45 % (40 C to 50 C)

Hide quick links: