Detailed Specifications

For all detailed specifications:
Accuracy is given as ([% of reading] + [number of least significant digits]) at 18 ºC to 28 ºC, with relative humidity up to
90 %, for a period of one year after calibration.
In the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100
% of range. The Meter is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add
-(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.

AC Voltage Function

Function
Range
Resolution
K
600.0 mV
0.1 mV
[2,4]
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
Low pass filter
[1]
Below 10 % of range, add 12 counts.
[2]
The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between
1 and 30 counts. A 30 count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce
a much larger constant error in later measurements.
[3]
Frequency range: 1 kHz to 2.5 kHz.
[4]
A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
[5]
Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.
46
45 – 65 Hz
30 - 200 Hz 200 - 440 Hz
± (0.7 % + 4)
± (0.7 % + 2)
± (0.7 % + 2) ± (1.0 % + 4)
Accuracy
440 Hz - 1 kHz
± (1.0 % + 4)
+1 % + 4
unspecified
[5]
-6 % - 4
1 - 5 kHz
5 - 20 kHz
± (2.0 % + 4)
± (2.0 % + 20)
[3]
± (2.0 % + 4)
unspecified
unspecified
unspecified
unspecified
unspecified