Specifications; General Specifications; Detailed Specifications - Fluke 87 V/AN Service Manual

Table of Contents

Advertisement

Specifications

General Specifications

Maximum Voltage between any Terminal and Earth Ground: 1000 V rms
Fuse Protection for mA or µA inputs: 44/100 A, 1000 V FAST Fuse
W
W
Fuse Protection for A input: 11 A, 1000 V FAST Fuse
Display: Digital: 6000 counts updates 4/sec. The Meter also has 19,999 counts in high-resolution mode.
Analog Bargraph: 33 segments, updates 40/sec. Frequency: 19,999 counts, updates 3/sec at > 10 Hz.
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C
Altitude: Operating: 2000 m; Storage: 10,000 m
Temperature Coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C)
Electromagnetic Compatibility: All ranges unless otherwise noted: In an RF field of 3 V/m total
accuracy = specified accuracy + 20 counts
Except: Temperature not specified.
Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 70% (35 °C to 55 °C)
Battery Type: 9 V zinc, NEDA 1604 or 6F22 or 006P
Battery Life: 400 hrs typical with alkaline (with backlight off)
Vibration: Per MIL-PRF-28800 for a Class 2 instrument
Shock: 1 Meter drop per IEC 61010-1:2001
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight: 12.5 oz (355 g)
Weight with Holster and Flex-Stand: 22.0 oz (624 g)
Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage
Category III, IEC 664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to
EN61010-1.

Detailed Specifications

For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18 °C to
28 °C, with relative humidity up to 90 %, for a period of one year after calibration
adjustment. In the 4 ½-digit mode, multiply the number of least significant digits
(counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range.
The Meter is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half
scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest
factor up to 3.
Tables 2 through 10 list the Meter's detailed specifications.
Digital Multimeter