Detailed Specifications; Dc Voltage, Conductance, And Resistance - Fluke 87V MAX Calibration

Hide thumbs Also See for 87V MAX:
Table of Contents

Advertisement

Battery Type .................................................................. 3 AA Alkaline batteries, NEDA 15A IEC LR6
Battery Life .................................................................... 800 hr typical without backlight (Alkaline)
Vibration ........................................................................ Per MIL-PRF-28800 for a Class 2 instrument
Size (H x W x L) ............................................................. 1.8 in x 3.7 in x 7.7 in (4.6 cm x 9.4 cm x 19.7 cm)
Size with Holster ........................................................... 2.4 in x 4.3 in x 8.5 in (6.0 cm x 10.1 cm x 21.5 cm)
Weight ............................................................................ 1.14 lb (517.1 g)
Weight with Holster and Flex-Stand............................ 1.54 lb (698.5 g)
IP Rating ........................................................................ IEC 60529: IP67

Detailed Specifications

For all detailed specifications:
Accuracy is specified for 2 years after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at
0 % to 95 %. Accuracy specifications take the form of ±([% of Reading] + [Number of least-significant digits]).
For the 4 ½-digit mode, multiply the number of least-significant digits (counts) by 10.
AC Voltage
AC conversions are ac-coupled and valid from 3 % to 100 % of range.
Range
Resolution
45 – 65 Hz
600.0 mV
0.1 mV
±(0.7 % + 4)
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
±(0.7 % + 2)
1000 V
1 V
Low-Pass Filter
[1] Below 30 Hz, use smoothing function. Below 20 Hz add 0.6 %.
[2] Below 10 % of range, add 12 counts.
[3] Frequency range: 1 to 2.5 kHz
[4] Specification increases from -1 % to -6 % at 440 Hz when filter is used.

DC Voltage, Conductance, and Resistance

Function
Range
mV dc
600.0 mV
6.000 V
60.00 V
V dc
600.0 V
1000 V
600.0 Ω
6.000 kΩ
60.00 kΩ
Ω
600.0 kΩ
6.000 MΩ
50.00 MΩ
60.00 nS
nS
[1] Add 0.5 % of reading when measuring above 30 MΩ in the 50 MΩ range, and 20 counts below 33 nS in the 60 nS range.
[2] When using the rel function to compensate for offsets.
15 – 200 Hz
200 – 440 Hz
±(1.0 % + 4)
+1.0 % + 4
[1]
±(1.0 % + 4)
[4]
-6.0 % - 4
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.01 nS
Accuracy
440 Hz – 1 kHz
1 – 5 kHz
±(2 % + 4)
[1]
±(2 % + 4)
Unspecified
Unspecified
Unspecified
Accuracy
±(0.1 % + 1)
±(0.05 % + 1)
±(0.2 % + 2)
±(0.2 % + 1)
±(1.0 % + 1)
±(1.0 % + 10)
Digital Multimeter
Detailed Specifications
5 – 20 kHz
[2]
±(2 % + 20)
Unspecified
[3]
Unspecified
Unspecified
Unspecified
[2]
[1]
[1,2]
3