Current Function - Fluke 87V Ex User Manual

True-rms multimeter
Hide thumbs Also See for 87V Ex:
Table of Contents

Advertisement

Current Function

Function
mA
60.00 mA
400.0 mA
\
6.000 A
(45 Hz to 2 kHz)
10.00 A
mA
60.00 mA
[
400.0 mA
6.000 A
10.00 A
600.0 µA
A B
6000 µA
(45 Hz to 2 kHz)
600.0 µA
AF
6000 µA
[1]
AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range (5 % to 100 % of range) and 10 A
range (15 % to 100 % or range).
[2]
The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading
between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of range. Using REL to offset this reading
may produce a much larger constant error in later measurements
[3]
W 10 A continuous up to 35 C; < 20 minutes on, 5 minutes off at 35 C to 55 C. 20 A for 30 seconds maximum; > 10 A unspecified.
[4]
400 mA continuous; 600 mA for 18 hrs maximum.
Range
Resolution
0.01 mA
[4]
0.1 mA
0.001 A
[3]
0.01 A
0.01 mA
[4]
0.1 mA
0.001 A
[3]
0.01 A
0.1 µA
1 µA
0.1 µA
1 µA
[1, 2]
Accuracy
Burden Voltage (typical)
±(1.0 % + 2)
1.8 mV/mA
± (1.0 % + 2)
1.8 mV/mA
± (1.0 % + 2)
0.03 V/A
± (1.0 % + 2)
0.03 V/A
± (0.2 % + 4)
1.8 mV/mA
± (0.2 % + 2)
1.8 mV/mA
± (0.2 % + 4)
0.03 V/A
± (0.2 % + 2)
0.03 V/A
± (1.0 % + 2)
100 µV/µA
± (1.0 % + 2)
100 µV/µA
± (0.2 % + 4)
100 µV/µA
± (0.2 % + 2)
100 µV/µA
True-rms Multimeter
Detailed Specifications
49