Output Performance Verification; Table 7-4 Output Performance Verification Tests - Agilent Technologies U1401A User's Manual And Service Manual

Handheld multi-function calibrator/meter
Hide thumbs Also See for U1401A:
Table of Contents

Advertisement

Output performance verification

Table 7-4
Output performance verification tests
Step
Function
1
Turn the rotary switch to
any one of the
positions
.
2
Turn the rotary switch to
any one of the
positions
.
3
Turn the rotary switch to
any one of the
positions.
U1401A User's and Service Guide
To verify the output functions of the Handheld
Multi- Function Calibrator/Meter, perform the verifications
tests listed in
Table 7-
the recommended test equipment for verifying each function.
Recommended test
equipment and connection
Connect the U1401A output
terminals to the 34405A
Multimeter.
Connect the U1401A output
terminals to the 34405A
Multimeter.
Connect the U1401A output
terminals to the 53131A
Universal Counter.
Performance Tests and Calibration
4. Refer to
Table 7- 1
Range or
U1401A
parameter
output
±1.5000 V
–1.5 V
0 V
+1.5 V
±15.000 V
–15 V
0 V
+15 V
±25.000 mA
–25 mA
+25 mA
Frequency
4.8 kHz
(10 kHz)
Frequency
600 Hz
(1 kHz)
Duty cycle
5 V, 25%
(0.39% to
@150 Hz
99.60%)
5 V, 75%@
150 Hz
7
on page 131 for
Error from
nominal 1
year
±0.75 mV
±0.3 mV
±0.75 mV
±7.5 mV
±3 mV
±7.5 mV
±0.125 µA
±0.125 µA
±0.25 Hz
±0.04 Hz
±0.203%
±0.208%
137