Agilent Technologies U1242A User's Manual And Service Manual page 66

Handheld digital multimeters
Hide thumbs Also See for U1242A:
Table of Contents

Advertisement

5
Performance Tests and Calibration
9 Turn the rotary switch to the next function according to the Test
10 Verify the adjustments using the
Table 5-15 Calibration Adjustments
Step
Test Function
1
Turn the rotary switch to
position
2
Press
to go to
function
52
Function column shown in
adjustment point shown in the calibration adjustment, see
Cal Range
V
Short
1000 mV
10 V
100 V
1000 V
V
1000 mV
10 V
100 V
1000 V
Table 5-
14. Repeat steps 3 to 8 for each
Performance Verification
Input Reference Value
Dual banana plug with
copper wires short
between two terminals
1 V
10 V
100 V
1000 V
30 mV, 70 Hz
1000 mV, 70 Hz
1000 mV, 1 kHz
1 V, 70 Hz
10 V, 70 Hz
10 V, 1 kHz
10 V, 70 Hz
100 V, 70 Hz
100 V, 1 kHz
100 V, 70 Hz
1000 V, 70 Hz
1000V, 1 kHz
U1241A/U1242A User's and Service Guide
Table 5-
15.
Tests.
Cal Item
U1241A
U1242A
SHrt
1000.0 mV
10.000 V
100.00 V
1000.0 V
30.0 mV
1000.0 mV
1000.0 mV
1.000 V
10.000 V
10.000 V
10.00 V
100.00 V
100.00 V
100.0 V
1000.0 V
1000.0 V

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1241a

Table of Contents