Step 3
Match the code on the meter with the code on the test strip vial
If the code on the meter does not match the code on the test strip
vial, press ▲ or ▼ to match the code number on the test strip vial.
The new code number will flash on the display for three seconds,
and then stay constant for three seconds.
The display will advance to the screen with the flashing blood drop
R
icon
.
If the codes already match, wait three seconds. The
display will advance to the screen with the flashing
R
blood drop icon
.
The meter is now ready to perform a blood glucose test.
NOTE:
• If the screen with the flashing blood drop icon
match, remove the test strip, wait until the meter turns off, then re-start from step 1, see
page 7.
• If you press ▲ by mistake so that the control solution test symbol CtL appears on the
display, press ▲ again to change it back to the screen with the flashing blood drop icon
æ
CAUTION: Matching the code on the meter and the code on the test strip vial is essential to
obtain accurate results. Each time you test, check to make sure the code numbers match.