Waters 2475 Multi-λ Flourescence Detector
Serial Number:
Checksum:
Version Number:
Date Installed:
Installed By:

Advertisement

loading
Need help?

Need help?

Do you have a question about the 2475 and is the answer not in the manual?

Questions and answers

Summary of Contents for Waters 2475

  • Page 1 Quick Install Guide Waters 2475 Multi-λ Flourescence Detector Serial Number: Checksum: Version Number: Date Installed: Installed By:...
  • Page 2 Connecting your detector Related documentation: Waters 2475 Multi- λ Fluorescence Detector Operator’s Guide, (71500247502, rD) Waters Ethernet Instrument Getting Started Guide, (7150074403, rB) Electrical Connections A. Connect the event or signal cables from the Detector Out 1 + Inject Start +...
  • Page 3 Column Outlet Line CAUTION: To prevent back-pressure from damaging the flow cell, always place the 2475 last in your HPLC system. A.Place a compression fitting and ferrule on the longer piece of stainless steel tubing (marked INLET), and connect the tubing to the column, detector, injector, or pump.
  • Page 4 Milford, MA 01757 508 478-2000 http://www.waters.com Waters is a registered trademark of Waters Corporation Printed in U.S.A. February, 2008 715000362, Rev. D If you have any questions regarding this installation, call your local Waters office. In the U.S. call 1-800-252-HPLC.