Keysight 85032F User's And Service Manual
Keysight 85032F User's And Service Manual

Keysight 85032F User's And Service Manual

Type-n 50 calibration kit

Advertisement

Quick Links

Keysight 85032F 
Type-N 50  Calibration Kit
User's and Service
Guide

Advertisement

loading
Need help?

Need help?

Do you have a question about the 85032F and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight 85032F

  • Page 1 Keysight 85032F  Type-N 50  Calibration Kit User’s and Service Guide...
  • Page 3 User’s and Service Guide Keysight Technologies 85032F Type-N 50 Calibration Kit Keysight Part Number: 85032-90021 Printed in USA Print Date: August 2010 Supersedes: December 2007 © Copyright 2000, 2002, 2004, 2007, 2010 Keysight Technologies. All rights reserved.
  • Page 4 REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
  • Page 5 Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Keysight web site: • Go to www.keysight.com. • Enter the document’s part number (located on the title page) in the Search box.
  • Page 6 85032F...
  • Page 7 How Keysight Verifies the Devices in This Kit ....... . .
  • Page 8 Returning a Kit or Device to Keysight ........
  • Page 9 General Information...
  • Page 10 Calibration Kit Overview Calibration Kit Overview The Keysight 85032F type-N 50calibration kit is used to calibrate Keysight network analyzer systems for measurements of components with type-N connectors up to 9 GHz. The standards in this calibration kit allow you to perform simple 1- or 2-port calibrations.
  • Page 11 If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions. Options There are several optional adapters available for the 85032F. See Table 6-1 on page 6-3 for further information.
  • Page 12 Table 6-1 on page 6-3 to verify a complete shipment. Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, contact Keysight Technologies. See Contacting Keysight on page 5-4.
  • Page 13 If you can not obtain repeatable measurements from your calibration standards, maintenance needs to be performed on the test port connectors, cables and calibration standards. Also, maintain at least one sample of the device under test or some known device as 85032F...
  • Page 14 Periodically remeasure the device and note any changes in its corrected response which can be attributed to the test system. With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again. 85032F...
  • Page 15 Specifications...
  • Page 16 During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in  Table 2-1. IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a heat source. 85032F...
  • Page 17 They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
  • Page 18 When measuring pin depth, the measured value (resultant average of three or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-3 to evaluate the condition of device connectors. 85032F...
  • Page 19 Definitions Values are Available on the Web Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 20 Specifications Electrical Specifications 85032F...
  • Page 21 Use, Maintenance, and Care of the Devices...
  • Page 22 3. Connect the other end of the cable to the test port. 4. Remove the short from the cable. Refer to Chapter 6, Replaceable Parts for ordering ESD protection supplies. Figure 3-1 ESD Protection Using Mat, Wrist Strap, and Grounded Power Cord 85032F...
  • Page 23 Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance. If a 85032F...
  • Page 24 These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will not make good electrical contact and must be replaced. NOTE This is particularly important when you are mating nonprecision to precision devices. 85032F...
  • Page 25 It may be harmful if absorbed through the skin. Wash thoroughly after handling.  In case of spill, soak up with sand or earth. Flush spill area with water.  Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulation. 85032F...
  • Page 26 Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector be you reassemble or use it. 4. Inspect Each Connector Inspect the connector again to make sure that no particles or residue are present. 85032F...
  • Page 27 Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only (see Table 6-2 on page 6-4 for part number information). They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for precise pin depth measurements.
  • Page 28 If the center conductor protrudes, the indicator will move clockwise to measure the amount of protrusion, which is read as a positive value. Figure 3-2 Reading the Connector Gage 85032F...
  • Page 29 Compare the average reading with the pin depth specifications listed in Table 2-3 on page 2-4. f. Without turning either the gage or the device, remove the device from the gage. 85032F...
  • Page 30 Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-3 Gaging Type-N Connectors 3-10 85032F...
  • Page 31 A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough. 8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables. 85032F 3-11...
  • Page 32 (greater than 90 degrees apart) will cause a lifting action that can misalign and stress the connections of the device involved This is especially true when several devices are connected together. Figure 3-4. Figure 3-4 Wrench Positions 3-12 85032F...
  • Page 33 1. Use an open-end wrench to prevent the device body from turning. 2. Use the torque wrench to loosen the connector nut. 3. Complete the separation by hand, turning only the connector nut. 4. Pull the connectors straight apart without twisting, rocking, or bending. 85032F 3-13...
  • Page 34 • Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-14 85032F...
  • Page 35 Performance Verification...
  • Page 36 Introduction Introduction The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment and calibration standards required to verify the specifications of the devices inside the kit have been specially manufactured and are not commercially available.
  • Page 37 A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Keysight Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Keysight Technologies. See Contacting Keysight on page 5-4.
  • Page 38 Performance Verification Recertification 85032F...
  • Page 39 Troubleshooting...
  • Page 40 Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. Figure 5-1 Troubleshooting Flowchart 85032F...
  • Page 41 5-4. Returning a Kit or Device to Keysight If your kit or device requires service, contact Keysight Technologies for information on where to send it. See Contacting Keysight on page 5-4. Include a service tag (found at the end of this manual) on which you provide the following information: •...
  • Page 42 Troubleshooting Contacting Keysight Contacting Keysight Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer.
  • Page 43 Replaceable Parts...
  • Page 44 Introduction Table 6-1 lists the replacement part numbers for the contents of the 85032F calibration kit. To order a listed part, note the description, the Keysight part number, and the quantity desired. Telephone or send your order to Keysight Technologies- see Contacting Keysight on page 5-4.
  • Page 45 Replaceable Parts Introduction Table 6-1 Replaceable Parts Item No. Description Keysight Part Number Calibration Devices (Type-N 50 ) Male offset open –m– 85032-60013 Female offset open 85032-60014 Female offset short 85032-60015 Male offset short –m– 85032-60016 Male broadband load –m–...
  • Page 46 Replaceable Parts Introduction Table 6-2 Items Not Included in Kit Description Keysight Part Number Miscellaneous Type-N gage set 85054-60049 Adapter Kit (type-N 50  to 3.5 mm) 11878A Wrenches 3/4 in, 135 N-cm (12 in-lb) torque wrench 8710-1766 1/2 in and 9/16 in open-end wrench...
  • Page 47 Standard Definitions...
  • Page 48 Keysight’s Calibration Kit Definitions Web page at http://na.support.keysight.com/pna/caldefs/stddefs.html. For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying Calibration Standards and Kits for Keysight Vector Network Analyzers.” This application note covers calibration standard definitions, calibration kit content and its structure requirements for Keysight vector network analyzers.
  • Page 49 Keysight Web site frequency of calibration cleaning connectors offsets, opens, shorts compressed air or nitrogen optional adapters connections 3-11 3-12...
  • Page 50 Keysight Web site static discharge storage 3-14 temperature affect on electrical performance calibration cautions about changes in device error-corrected measurement specifications...
  • Page 51 This information is subject to change without notice. © Keysight Technologies 2000 - 2014 August 2014 *85032-90021* 85032-90021 www.keysight.com...