7 mm Calibration Kit User’s and Service Guide General Information Calibration Kit Overview The Keysight 85031B 7 mm calibration kit is used to calibrate Keysight network analyzers up to 6 GHz for measurements of components with 7 mm connectors. Kit Contents The 85031B calibration kit includes the following items: —...
— entered from the front panel Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at https://www.keysight.com/us/en/assets/9922-01521/technical-specificatio ns/Calibration-Kit-Definitions.pdf.
Keysight will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. When you send the kit or device to Keysight, include a service tag (found after the index of this manual) with the following information: —...
With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again. Keysight 85031B User’s and Service Guide...
Keysight 7 mm Calibration Kit User’s and Service Guide Specifications Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits Temperature +15 °C to +35 °C Operating Storage −40 °C to +75 °C Error-corrected range ± 1 °C of measurement calibration temperature Relative humidity Type tested, 0% to 95% at 40 °C, non-condensing...
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
Page 17
+0.0051 to -0.0635 Combination open/short Open: 0 to -0.001 0 to -0.0254 Short: +0.0001 to -0.0001 +0.00254 to -0.00254 Center conductor recession with collet removed = 0.000 to 0.002 inches. Figure 2-1 Connector Pin Depth Keysight 85031B User’s and Service Guide...
Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other...
Keysight 7 mm Calibration Kit User’s and Service Guide Use, Maintenance, and Care of the Devices Electrostatic Discharge Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Page 20
Use, Maintenance, and Care of the Devices Electrostatic Discharge Figure 3-1 ESD Protection Setup Keysight 85031B User’s and Service Guide...
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance. Keysight 85031B User’s and Service Guide...
Page 22
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. Keysight 85031B User’s and Service Guide...
Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. Keysight 85031B User’s and Service Guide...
Refer to the calibration kit documentation for detailed connector care information. For course numbers about additional connector care instruction, contact Keysight Technologies. Refer to “Contacting Keysight” on page 5-4. Keysight 85031B User’s and Service Guide...
Page 25
Cleaning connectors with alcohol shall only be done with the instrument’s power cord removed, and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument. Keysight 85031B User’s and Service Guide...
Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. See Table 6-2 on page 6-2 for part number information. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.
Refer to “Pin Depth” on page 2-2 for definitions of protrusion and recession. Figure 3-2 Reading the Connector Gage Keysight 85031B User’s and Service Guide...
Using an open-end wrench to keep the gage from rotating, use the torque wrench recommended for use with the kit to tighten the device connector nut to the specified torque. Refer to “Final Connection Using a Torque Wrench” on page 3-15 for more information. 3-10 Keysight 85031B User’s and Service Guide...
Page 29
Read the gage indicator dial. If the needle has moved clockwise, the center conductor is protruding by and amount indicated by the black numbers. If the needle has moved counterclockwise, the center conductor is recessed by an amount indicated by the red numbers. Keysight 85031B User’s and Service Guide 3-11...
Page 30
Compare the average reading with the pin depth specifications listed in Table 2-2 on page 2-3. g. Remove the device from the gage and replace the center conductor collet. 3-12 Keysight 85031B User’s and Service Guide...
Page 31
Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-3 Gaging 7 mm Connectors Keysight 85031B User’s and Service Guide 3-13...
Very light finger pressure is enough to accomplish this. 10.Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables. 3-14 Keysight 85031B User’s and Service Guide...
Page 33
This is especially true when several devices are connected together. Figure 3-4 Wrench Positions 2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-5. Keysight 85031B User’s and Service Guide 3-15...
2. Use the torque wrench to loosen the connector nut. 3. Complete the separation by hand, turning only the connector nut. 4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors. 3-16 Keysight 85031B User’s and Service Guide...
— Do not store connectors loose in a box, or in a desk or bench drawer. This is the most common cause of connector damage during storage. Keysight 85031B User’s and Service Guide 3-17...
Page 36
Use, Maintenance, and Care of the Devices Handling and Storage 3-18 Keysight 85031B User’s and Service Guide...
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies.
The recertification interval should begin on the date the kit is first used after the recertification date. Where to Send a Kit for Recertification Contact Keysight for information on where to send your kit for recertification. For contact information, see “Contacting Keysight” on page 5-5.
Keysight 7 mm Calibration Kit User’s and Service Guide Troubleshooting Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
Page 40
Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart Keysight 85031B User’s and Service Guide...
2. Enter your VNA model number (Ex: 8753ES) in the Search box and click Search. 3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF you want to view. If you need additional information, see “Contacting Keysight” on page 5-5. Keysight 85031B User’s and Service Guide...
Troubleshooting Returning a Kit or Device to Keysight Returning a Kit or Device to Keysight If your kit or device requires service, contact Keysight Technologies for information on where to send it. See “Contacting Keysight” on page 5-5 contact information. Include a service tag (located near the end of this manual) on which you provide the following information: —...
Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer. In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number.
Page 44
Troubleshooting Contacting Keysight Keysight 85031B User’s and Service Guide...
7 mm Calibration Kit User’s and Service Guide Replaceable Parts Introduction Table 6-1 lists the replacement part numbers for items included in the 85031B calibration kit and Figure 6-1 illustrates each of these items. Table 6-2 lists the replacement part numbers for items not included in the calibration kit that are either required or recommended for successful operation of the kit.
Page 46
2 ft by 4 ft conductive table mat with 15 ft grounding wire 9300-0797 ESD heel strap 9300-1308 Connector Cleaning Supplies Anhydrous isopropyl alcohol (>92% pure) Foam-tipped cleaning swabs 9301-1745 a. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally. Keysight 85031B User’s and Service Guide...
Calibration Kit Definitions Web page at https://www.keysight.com/us/en/assets/9922-01521/technical-specificatio ns/Calibration-Kit-Definitions.pdf. For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying Calibration Standards and Kits for Keysight Vector Network Analyzers.” This application note covers calibration standard definitions, calibration kit content and its structure requirements for Keysight vector network analyzers.
Page 48
Standard Definitions Class Assignments and Standard Definitions Values are Available on the Web Keysight 85031B User’s and Service Guide...
Page 49
4–1 disconnecting 3–16 cleaning 3–5 care of connectors 3–6 handling 3–17 inspecting 3–3 center conductor maintenance 1–4 visual inspection protrusion 3–3 3–9 performance wear 3–3 recession 3–9 verifying 4–1 affect electrical Keysight 85031B User’s and Service Guide I–1...
Page 50
3–8 inspection of 3–3 recession 2–2 male connectors mechanical characteristics 3–10 2–2 typical values 2–3 to determine pin depth affect on electrical 3–8 preventive maintenance 1–4 performance 2–2 protrusion verifying 3–8 I–2 Keysight 85050B User’s and Service Guide...
Page 51
, 4–1, environmental 2–1 4–2 frequency 2–4 troubleshooting 5–1 humidity 2–1 return loss 2–4 temperature 2–1 torque wrench 3–15 verification traceability , 4–1, 4–2 temperature 2–1 standard definitions visual inspection 3–3 Keysight 85031B User’s and Service Guide I–3...
Page 52
Index I–4 Keysight 85031B User’s and Service Guide...