Figure 5–26. Configuration Settings
From pH (or ORP) measurement mode press ENT key to
1 1
enter setup mode as described in page 24. The LCD shows the
first screen of setup mode (OFS). Press ▲ or ▼ key to select
configuration settings screen (Cnfg).
Press ENT key to access configuration settings (Cnfg).
Enable/disable LCD backlight: The upper display shows
2
'LitE'. The lower display shows last configured backlight
selection ('On' or 'OFF').
Press ▲ or ▼ key to enable (LitE On) or disable (LitE OFF)
backlight of the LCD.
Press ENT key to confirm your selection.
Select measurement mode: The upper display shows the last
3
configured measurement mode ('PH' or 'OrP').
Press ▲ or ▼ key to select your desired measurement mode.
Press ENT key to confirm your selection.
NOTE: When the measurement mode is changed, the relay
operation parameters (described in page 28) will be reset back
33