Security Codes

This controller has two levels of security protection with separate security codes. The first level allows
entry into the Calibration mode: security code = 11; the second allows entry into the SETUP mode:
security code = 22.
The security codes protect the controller from unauthorised tampering of its current setting. The
parameters cannot be changed unless the security code is entered.
5.2.1

How to enter and change parameters in Calibration mode

1) Press the CAL key. The upper display shows "000" and the
lower display shows "C.Cd" to prompt the user to enter the
Calibration security code.
2) Press the ▲ or ▼ keys to scroll upper display to Calibration
security code "11".
3) Press the ENTER key. The display shows "CAL CON".
4) Press ENTER key again to begin calibration. Refer to Section 6 for full details on calibration.
5) Press the ▲ or ▼ keys simultaneously (escape) to return to the Measurement mode.
NOTE: To view (not change) the SETUP parameters, push the ENTER key when the security
code reads "000".
5.2.1.1
Clearing the Calibration security code from the display
The calibration security code automatically resets from "11" to "000" after you return to Measurement
mode, so you do not need to clear the security code from the display.
5.2.2

How to enter and change parameters in Advanced Setup mode

1) Press the ENTER key once. The upper display shows "000" and the lower display shows "S.Cd"
to prompt you to enter the Advanced Setup security code.
2) Press the ▲ or ▼ keys to scroll the display to Setup security code "22". NOTE: Pressing the
ENTER key at a value other than "22" causes the controller to revert to the Measurement mode.
3) Press the ENTER key.
4) The upper display reads "tc".
5) You are now in the Advanced Setup mode. See Section 7 for complete instructions. To return to
Measurement mode, press the ▲ or ▼ keys simultaneously (escape).
NOTE: If you want to view (not change) set up parameters, push the ENTER key when the
security code reads "000".